Z-1322-2014Class II

Integrity 1.1. Integrity is the interface and control software for the Elekta range of medical digital linear...

Elekta Limited / initiated 2013-03-28 / Terminated / root cause: software

Elekta, Inc. began a recall of Integrity 1.1. Integrity is the interface and control software for the Elekta range of medical digital linear accelerators and is intended to assist a licensed... on . FDA classified it as Class II and gives the reason as "The problem is that the "static tolerances" from the calibration files (loaded database) are used for beam delivery instead of the machine calculated values". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1322-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Integrity 1.1. Integrity is the interface and control software for the Elekta range of medical digital linear accelerators and is intended to assist a licensed practitioner in the delivery of radiation to defined target volumes.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
The problem is that the "static tolerances" from the calibration files (loaded database) are used for beam delivery instead of the machine calculated values.
Root cause tag
software
FDA root cause
Software design
Action
An Important Field Safety Notice (IFSN) has been released and distributed to all customers. Integrity 1.1 Incorrect Static Tolerances for Linac Geometric Axes, 200 01 502 032. The notice explains the issue and that the user should apply the static Tolerance Updater tool to check tolerances and adjust as required. Important Field Safety Modification Digital Accelerator FCO Ref: 200 02 502 031 Integrity R1.1 Static Tolerance Updater Tool was released 03-2013. Elekta has released Integrity version 1.2 which contains the permanent fix for this issue. The permanent solution fixes the error that made the digital accelerator protection functions use the value loaded from the calibration database. The solution now ensures the protection functions use the values the digital accelerator calculates. Important Field Safety Notice 200 01 502 035 has been released to inform customers the solution is available. Important Field Safety Modification 200 02 502 036 will track the installation of the software by Elekta service engineers.
Quantity in commerce
498
Distribution
Worldwide Distribution-USA (nationwide) including the states of AL, AZ, CT, FL, GA, KY, LA, MD, MO, NM, NY, NC, OH, OK, PA, TN, TX, and WA and the countries of Argentina, Australia, Austria, Belarus, Belgium, Brazil, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, France, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Iraq, Ireland, Italy, Japan, Jordan, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Poland, Romania, Russia, Saudi Arabia, Senegal, Singapore, Slovakia, South Korea, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.
Product codes and lots
152718, 152003, 152007, 151963, 152081, 151693, 152131, 152318, 105792, 105570, 152778, 151052, 152363, 152472, 152195, 153022, 152689, 152640, 152642, 152663, 152680, 152782, 152796, 152829, 152862, 152962, 153030, 153125, 153244, 153270, 153211, 153296, 152834, 152596, 151354, 151353, 152664, 151200, 151823, 151822, 151610, 152026, 152027, 152692, 151276, 152856, 151106, 152877, 152986, 152810, 105935, 153105, 153069, 151023, 152914, 151975, 151240, 151239, 153135, 105810, 153145, 153129, 151182, 151650, 152886, 151850, 152831, 153132, 151343, 153063, 105985, 152832, 152845, 152837, 152863, 152859, 152776, 152857, 152866, 152876, 152820, 152870, 152884, 152865, 152872, 152909, 152910, 152911, 152918, 152934, 152961, 152966, 153009, 153041, 153026, 153078, 153079, 153083, 153070, 153066, 153027, 153154, 153157, 153159, 153165, 153131, 153352, 152675, 152737, 152299, 152403, 152676, 151670, 152252, 151153, 151482, 151296, 152068, 152286, 152097, 151554, 151768, 151853, 151854, 151786, 151495, 151787, 153099, 152935, 151865, 151466, 105920, 152521, 151536, 151263, 105988, 151028, 151058, 152719, 151174, 151597, 152645, 152650, 152647, 105755, 152743, 151857, 151921, 105896, 151986, 152855, 152092, 151006, 151277, 105314, 105919, 152746, 152770, 105901, 151635, 152575, 152627, 151701, 151922, 151852, 152422, 152309, 152498, 152628, 152706, 152616, 105850, 152014, 151688, 151638, 152721, 153234, 152816, 105554, 105987, 151452, 151166, 152913, 151740, 152042, 152904, 152254, 152753, 152554, 152591, 152644, 152607, 152648, 152687, 152714, 152899, 152739, 152672, 152900, 152952, 152977, 153080, 153169, 153183, 152908, 153349, 153325, 152702, 152869, 152724, 151529, 152840, 152841, 151442, 151445, 151612, 151613, 151756, 105677, 151755, 151189, 151034, 151652, 152757, 105893, 105892, 152945, 151060, 105711, 105997. 151324, 151316, 151318, 151385, 152873, 151660, 152734, 152646, 152682, 151087, 151088, 105833, 152613, 152958, 105946, 151117, 151308, 151108, 151280, 151912, 152106, 152157, 152294, 152637, 152624, 152667, 152727, 152726, 152750, 152893, 152953, 151132, 151150, 151879, 151583, 152595, 152733, 153087, 152975, 153310, 153229, 152738, 152735, 152785, 152797, 152947, 153327, 152907, 153291, 153290, 152677, 152674, 151112, 105937, 151349, 105838, 105758, 105927, 151045, 151314, 151553, 151140, 151101, 105995, 105915, 151405, 152638, 152325, 152457, 152512, 151196, 151368, 152118, 152119, 105968, 151040, 105912, 151247, 105943, 151005, 105948, 151341, 151898, 151703, 151982, 152161, 151718, 153059, 151297, 151278, 151020, 105928, 152603, 153084, 152942, 151829, 151284, 152298, 151449, 151144, 151602, 152708, 152713, 152605, 152606, 152614, 152611, 152629, 152679, 152799, 152880, 152928, 152920, 152336, 153343, 153348, 152101, 151669, 151237, 151121, 153205, 151897, 105415, 152793, 105640, 105703, 105869, 151175, 151066, 151233, 152060, 151503, 152190, 153333, 152795, 151089, 152660, 152538, 152827, 152848, 152442, 152846, 152723, 152921, 151119, 152769, 152968, 152143, 152720, 151497, 151197, 151420, 105721, 151079, 105787, 105685, 151511, 151022, 105743, 105742, 151249, 151042, 105786, 151516, 151109, 151331, 151524, 152592, 151255, 105813, 151556, 151526, 105811, 151032, 151581, 151243, 151492, 152070, 151096, 151004, 151136, 105797, 105802, 105957, 105789, 151292, 151425, 151291, 152148, 151332, 151333, 151486, 151415, 151573, 151528, 151487, 151527, 151572, 151414, 105874, 151225, 152653, 105814, 151092, 151047, 151706, 151691, 151692, 151934, 151945, 151946, 153190, 151127, 105412, 152030, 152168, 152396, 152654, 151749, 152891, 152878, 152858, 105894, 151033, 151258, 151168, 151159, 151228, 152760, 151791, 152522, 151391, 151216, 151539, 151611, 151696, 151867, 152965, 151325, 152220, 151785, 152217, 151388, 151394, 151892, 152043, 151793, 151684, 152223, 152303, 152358, 151683, 152652, 152659, 152703, 151672, 152767, 152777, 152783, 152784, 152787, 152798, 152881, 151710, 153045, 152929, 153137.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1322-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1322-2014Class IITerminatedVoluntary: Firm initiated