Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1196-2014Class II

Gyrus ACMI Disposable Falope-Ring Band Applicator Kits. REF 006889-901 and REF 005280-901 (with 8 mm...

Gyrus ACMI, Inc. / initiated 2014-02-19 / Terminated / root cause: manufacturing process

Gyrus Medical, Inc began a recall of Gyrus ACMI Disposable Falope-Ring Band Applicator Kits. REF 006889-901 and REF 005280-901 (with 8 mm Disposable Trocan). Sterile EO. Single Use Only. The... on . FDA classified it as Class II and gives the reason as "Due to an anomaly in the packaging process for specific lots of Disposable Falope-Ring Band Kit, it is possible that a defective seal could be present that may allow a breach of the package's sterile barrier and may...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1196-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Gyrus ACMI, Inc.
Product
Gyrus ACMI Disposable Falope-Ring Band Applicator Kits. REF 006889-901 and REF 005280-901 (with 8 mm Disposable Trocan). Sterile EO. Single Use Only. The disposable Falope-Ring Band Applicator is a sterile, two-band, dual-incision instrument. The two-band applicator enables the physician to occlude both Fallopian tubes without removing the applicator from the abdominal cavity. The Falope-Ring Band and Applicator Kits are indicated for use for female sterilization (permanent contraception).
Product code
KNH Laparoscopic Contraceptive Tubal Occlusion Device
Reason for recall
Due to an anomaly in the packaging process for specific lots of Disposable Falope-Ring Band Kit, it is possible that a defective seal could be present that may allow a breach of the package's sterile barrier and may compromise the sterility of the product. The breach may or may not be easily seen.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Olympus sent an Urgent Medical Device Recall letter dated February 19, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Olympus requires you to take the following action: I. Immediately cease any further use of any affected product you have, remove it from your inventory and quarantine it until it is shipped back to us. 2. Call your Olympus customer service representative at 1-888-524-7266 to obtain a Returned Goods Authorization so that you may return the product with no charge to you. Olympus will issue a credit t or replacement to your facility y for any returned product. 3. Please note on the enclosed questionnaire that you have received this information. 4. Fax the completed questionnaire to 484-896-7128 regardless of whether you have any affected inventory at your facility. ln addition, if you may have further distributed this product, please identify your customers, notify them at once of this product recall, and appropriately document your notification process. Your notification to your customers may be enhanced by including a copy of this recall notification letter.
Quantity in commerce
856 units
Distribution
US Distribution.
Product codes and lots
Lots: MK761655, MK764022, MK767655, MK526401

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1196-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1196-2014Class IITerminatedVoluntary: Firm initiated