Z-1246-2014Class II
Oxoid Vancomycin 30 mcg, each cartridge contains 50 susceptibility discs, CT0058B. The firm name on the label...
Remel Inc / initiated 2014-02-18 / Terminated / root cause: manufacturing process
Remel Inc began a recall of Oxoid Vancomycin 30 mcg, each cartridge contains 50 susceptibility discs, CT0058B. The firm name on the label is Oxoid Ltd., Basingstoke, Hampshire, England... on . FDA classified it as Class II and gives the reason as "Some of the discs may not be impregnated with the antibiotic". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1246-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Remel Inc
- Product
- Oxoid Vancomycin 30 mcg, each cartridge contains 50 susceptibility discs, CT0058B. The firm name on the label is Oxoid Ltd., Basingstoke, Hampshire, England. Used in the semi-quantitative agar diffusion test method for in-vitro susceptibility testing.
- Product code
- JTN Susceptibility Test Discs, Antimicrobial
- Reason for recall
- Some of the discs may not be impregnated with the antibiotic.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- ThermoFisher sent an Urgent Medical Device Recall letter dated February 18, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to notify any personnel who need to be notified of the potential for a false indication of resistance to Vancomycin. Customers were also instructed to review their inventory for the affected product and discard it. Customers should review results reported using this lot and consider retesting and/or seek appropriate expert advice. Return the attached Medical Device Recall Return Response acknowledging receipt of the notice and disposal of the affected product. If product was further distributed, customers should contact those entities, advise them of the situation and provide them with a copy of the letter. Customers with questions were instructed to contact the Technical Services Department at 800-255-6730 (US) and 913-888-0939 (International). For questions regarding this recal lcall 913-895-4185.
- Quantity in commerce
- 40/5/50-disc cartridge packages
- Distribution
- Nationwide Distribution
- Product codes and lots
- Lot 1174497, exp. 2015/04
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1246-2014 | Class II | Terminated | Voluntary: Firm initiated |