Z-1482-2014Class I

Abbott Acclaim Infusion Pump Designed to deliver parenteral infusions including whole blood or red blood cell...

Abbott Medical / initiated 2013-10-31 / Terminated / root cause: design

Hospira Inc. began a recall of Abbott Acclaim Infusion Pump Designed to deliver parenteral infusions including whole blood or red blood cell components and enteral fluids using a wide... on . FDA classified it as Class I and gives the reason as "Hospira has received customer reports of broken door assemblies on the Hospira/Abbott Acclaim Encore infusion pumps. If the door assembly breaks, it may prevent the door from closing properly and unrestricted flow may...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1482-2014
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Abbott Acclaim Infusion Pump Designed to deliver parenteral infusions including whole blood or red blood cell components and enteral fluids using a wide variety of standard administration sets and fluid containers.
Product code
FRN Pump, Infusion
Reason for recall
Hospira has received customer reports of broken door assemblies on the Hospira/Abbott Acclaim Encore infusion pumps. If the door assembly breaks, it may prevent the door from closing properly and unrestricted flow may occur. If the door cannot be closed, the pump cannot be used which can result in a delay in therapy.
Root cause tag
design
FDA root cause
Component design/selection
Action
URGENT MEDICAL DEVICE CORRECTION letters, dated October 31, 2013, were sent to all direct accounts. The letters identified the affected product, as well as the reason for recall, risk to health, and instructions for device users. Users are to inspect each Hospira/Abbott Acclaim Encore infusion pump for door handle cracks prior to programming a therapy by following the provided instructions. The letter also included instructions for recipients to ensure that all potential users in the facility are made aware of the safety notification and the recommended actions; and complete the attached reply form and return it to the fax number or e-mail address on the form. Customers are to contact Stericycle at 866-891-0586 (Monday - Friday, 8:00 AM - 5:00 PM ET) to obtain additional reply forms if needed. If the devices were further distributed, those customers should be notified and asked to contact Stericycle at 866-891-0586 (Monday - Friday, 8:00 AM - 5:00 PM ET) to receive a reply form. The letter reminds customers that Hospira is retiring the Acclaim Encore infusion pump and will no longer support them as of December 31, 2014.
Quantity in commerce
13,218 infusion pumps total
Distribution
Nationwide Distribution -- AK, AL, AR, AZ, CA, CO, CR, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.
Product codes and lots
List Number: 12032;  Serial Numbers: 12322921, 12323859, 12322780, 12322801, 12322810, 12322812, 12322878, 12322880, 12322887, 12322896, 12322903, 12322904, 12322911, 12322917, 12322936, 12322939, 12323867, 12322916, 12323056, 12321139, 12324767, 12324854, 12324802, 12322873, 12324375, 1231453, 1231610.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1482-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1482-2014Class ITerminatedVoluntary: Firm initiated
Z-1483-2014Class ITerminatedVoluntary: Firm initiated