Z-1336-2014Class II
Playtex Nurser Deluxe Double Electric Breast Pump, Model X06578AO, powered by a 12V DC power supply, Model...
Energizer Personal Care / initiated 2014-03-18 / Terminated
Energizer Personal Care began a recall of Playtex Nurser Deluxe Double Electric Breast Pump, Model X06578AO, powered by a 12V DC power supply, Model SY-12120, which is included with the package on . FDA classified it as Class II and gives the reason as "Some of the power adapters outer casings may become loose and separate, resulting in a potential for electric shock". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1336-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Energizer Personal Care
- Product
- Playtex Nurser Deluxe Double Electric Breast Pump, Model X06578AO, powered by a 12V DC power supply, Model SY-12120, which is included with the package.
- Product code
- HGX Pump, Breast, Powered
- Reason for recall
- Some of the power adapters outer casings may become loose and separate, resulting in a potential for electric shock.
- FDA root cause
- Component change control
- Action
- Energizer sent and Voluntary Retail Recall Notice letter dated March 18, 2014 to retail customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to stop sale and return the affected product. The letter provides instruction to customers to return the affected product. For questions or concerns, contact Energizer Personal Care at 1-888-207-1492 or visit www.playtexproducts.com. On 9/9/2014 the firm expanded the recall to one additional production lot to include: Suspect date codes for Breast Pumps: P12315-XXXX to P13205-XXXX. Suspect adaptor date codes: 1238 to 1324 Revised Press issued 9/9/2014.
- Quantity in commerce
- 35,803
- Distribution
- USA Nationwide Distribution in the states of: AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, MD, MI, MN, MO, NC, NJ, NV, NY, OH, ON, OR, PA, SC, TN, TX, VA, VT, WA, WI Canada - one consignee, Energizer (recalling firm's) company store
- Product codes and lots
- Suspect date codes for Breast Pumps: P12324-91667C to P13205-30673C. Suspect adaptor date codes: 1241 to 1324. 9/9/2014 Expanded recall: Suspect date codes for Breast Pumps: P12315-XXXX to P13205-XXXX. Suspect adaptor date codes: 1238 to 1324.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1336-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY