Z-1219-2014Class II

Sarns Malleable Dual-stage Venous Return Cannulae, 28/38 Fr with 1/2" connector, 14.5" (37 cm) long, with X...

Terumo Cardiovascular Systems Corporation / initiated 2014-03-03 / Terminated / root cause: design

Terumo Cardiovascular Systems Corporation began a recall of Sarns Malleable Dual-stage Venous Return Cannulae, 28/38 Fr with 1/2" connector, 14.5" (37 cm) long, with X Coating The Sarns Malleable Dual-stage Venous... on . FDA classified it as Class II and gives the reason as "During in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1219-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Sarns Malleable Dual-stage Venous Return Cannulae, 28/38 Fr with 1/2" connector, 14.5" (37 cm) long, with X Coating The Sarns Malleable Dual-stage Venous Return Catheters and Sarns Dual-stage Venous Return Cannula are indicated for single-tube venous drainage from the right atrium and vena cava during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
Product code
DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Reason for recall
During in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return Cannulae.
Root cause tag
design
FDA root cause
Device Design
Action
Terumo sent an Urgent Medical Device Recall letter dated March 3, 2014, to all affected consignees. The letter included the reason for removal, potential hazards, and customer instructions. Terumo CVS advised users to stop using the affected devices and return any unused product on hand. Questions or concerns were directed to Terumo CVS Customer Service: 1-800-521-2818 M-F, 8am-6pm. For questions regarding this recall call 734-741-6173.
Distribution
Worldwide Distribution - USA including CA, CT, FL, GA, IL, IN, LA, MI, MN, MO, NC, NY, PA, TN, TX, and Internationallly to United Arab Emirates (UAE), Taiwan, Thailand, Malaysia, Belgium, Japan, Mexico, and Canada.
Product codes and lots
Catalog Number 4934X

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1219-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1218-2014Class IITerminatedVoluntary: Firm initiated
Z-1219-2014Class IITerminatedVoluntary: Firm initiated
Z-1220-2014Class IITerminatedVoluntary: Firm initiated
Z-1221-2014Class IITerminatedVoluntary: Firm initiated
Z-1222-2014Class IITerminatedVoluntary: Firm initiated
Z-1223-2014Class IITerminatedVoluntary: Firm initiated
Z-1224-2014Class IITerminatedVoluntary: Firm initiated