Z-1553-2014Class II

ADVIA Centaur XP Immunoassay system, an automated in vitro diagnostic analyzer. Catalog # 078-A011-03 - ADVIA...

Siemens Healthcare Diagnostics Inc. / initiated 2014-02-18 / Terminated / root cause: design

Siemens Healthcare Diagnostics began a recall of ADVIA Centaur XP Immunoassay system, an automated in vitro diagnostic analyzer. Catalog # 078-A011-03 - ADVIA Centaur XP, Catalog # 078-A011R03 - ADVIA Centaur... on . FDA classified it as Class II and gives the reason as "Instrument Cover Gas Spring failures. Over time, the gas spring may lose its effectiveness and fail to support the cover in its full or partially open position. This may lead to the cover falling". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1553-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ADVIA Centaur XP Immunoassay system, an automated in vitro diagnostic analyzer. Catalog # 078-A011-03 - ADVIA Centaur XP, Catalog # 078-A011R03 - ADVIA Centaur XP refurb.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Instrument Cover Gas Spring failures. Over time, the gas spring may lose its effectiveness and fail to support the cover in its full or partially open position. This may lead to the cover falling.
Root cause tag
design
FDA root cause
Device Design
Action
On February 18, 2014 Siemens Healthcare Diagnostics distributed Urgent Medical Device Correction notices to their customers via courier service (Urgent Medical Device Correction/). Upon receipt of the communication, customers are asked to follow "Actions to be Taken by the Customer". If the cover is unable to stay partially raised, immediately contact your Siemens technical support representative to arrange for a service. Complete and return the Field Correction Effectiveness Check attached to this letter within thirty (30) days by faxing to (302) 631-7597. Customers with questions should contact their Siemens technical support representative.
Quantity in commerce
5,900 units total (1,679 domestically & 4,221 internationally)
Distribution
Nationwide and internationally
Product codes and lots
All serial numbers of Catalog # 078-A011-03 - ADVIA Centaur XP and Catalog # 078-A011R03 - ADVIA Centaur XP refurb.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1553-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1553-2014Class IITerminatedVoluntary: Firm initiated