Z-1458-2014Class II
BIOMET 3i Endosseous Dental Implant
ZimVie US Corp LLC / initiated 2013-08-29 / Terminated / root cause: manufacturing process
Biomet 3i, LLC began a recall of BIOMET 3i Endosseous Dental Implant on . FDA classified it as Class II and gives the reason as "Due to residual machining fluid identified by discoloration (darker in appearance) on the external dental implant surface". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1458-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ZimVie US Corp LLC
- Product
- BIOMET 3i Endosseous Dental Implant.
- Product code
- DZE Implant, Endosseous, Root-Form
- Reason for recall
- Due to residual machining fluid identified by discoloration (darker in appearance) on the external dental implant surface.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Biomet 3i sent an Urgent Medical Device Notice dated September 9, 2013, to all affected customers. The notice identified the product, problem and actions to be taken. Each customer was instructed to check their respective inventory for the affected product and to return any unused quantities along with the attached Recall Response Form. Biomet 3i will issue replacement product at no charge. For questions or concerns, call Biomet 3i at 1-800-342-5454.
- Quantity in commerce
- 34,744
- Distribution
- Worldwide Distribution - USA and countries of United Arab Emirates, Argentina, Austria, Australia, Belgium, Brazil, Canada, Switzerland, Chile, Colombia, Costa Rica, Germany, Dominican Republic, Spain, France, Great Britain (UK), Greece, Hong Kong, Ireland, Israel, India, Italy, Japan, Korea (South), Lebanon, Liechtenstein, Lithuania, Mexico, Netherlands, Poland, Portugal, Paraguay, Saudi Arabia, Sweden, Singapore, Turkey, Taiwan, Uruguay, Vietnam and South Africa.
- Product codes and lots
- Model Number(s): Affected devices catalog names are: BNES310, BNET3210, BNPS6511, BNPS6513, BNPS6585, BNPT6510, BNPT6511, BNPT6513, BNPT6585, BNSS310, BNSS311, BNSS313, BNSS610, BNSS613, BNSS615, BNSS685, BNST3210, BNST3211, BNST3213, BNST3215, BNST3285, BNST611, BNST613, BNST685, BOPS6510, BOPS6511, BOPS6515, BOPS6585, BOPT6510, BOPT6511, BOPT6513, BOPT6585, BOSS310, BOSS311, BOSS313, BOSS315, BOSS611, BOSS613, BOSS685, BOST3210, BOST3211, BOST3213, BOST3215, BOST3285, BOST610, BOST611, BOST613, BOST685, FNT3210, FNT3211, IFNT3210, IFNT3211, IFNT3213, IFNT3215, IFNT3285, IFNT610, IFNT611, IFNT613, IFNT685, IFOSS5611, IFOS5685, IFOS610, IFOS611, IFOS615, IFOSM310, IFOSM311, IFOSM313, IFOSM385, IIOS5610, IIOS5611, INT3210, INT3211, INT3213, INT3215, INT3285, INT610, INT611, INT613, INT685, IOSM311, IOSM313, IOSM315, IOSM385, IOSS610, IOSS613, NIIITP6510, NIIITP6511, NIIITP6513, NIIITP6515, NIIITP6585, NINT3210, NINT3211, NINT3213, NINT3215, NINT610, NINT611, NINT613, NINT685, NIOSM311, NIOSM313, NIOSS610, NT3210, NT3211, OSM310, XIFNT3210, XIFNT3211, XIFNT3213, XIFNT3215, XIFNT3285, XIFN610, XIFNT611, XIFNT613, XIFNT685, XIFOSM310, XIFOSM311, XIFOSM313, XIFOSS610, XIFOSS611, XIIOS6510, XIIOS6511, XIIOS6513, XIITP6510, XIITP6511, XIITP6513. Manufactured dates from November 5, 2012 through August 21, 2013.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1458-2014 | Class II | Terminated | Voluntary: Firm initiated |