Z-1230-2014Class I

Heart Mate II Pocket Controllers removed from packaging

Abbott Medical / initiated 2014-03-04 / Terminated / root cause: design

Thoratec Corporation began a recall of Heart Mate II Pocket Controllers removed from packaging Product Usage: HeartMate II is a mechanical device that circulates blood throughout the body when the... on . FDA classified it as Class I and gives the reason as "Discovery of serious injuries and deaths associated with the process of changing from a primary System controller to their back-up System controller in patients using the Pocket System controller model". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1230-2014
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Heart Mate II Pocket Controllers removed from packaging Product Usage: HeartMate II is a mechanical device that circulates blood throughout the body when the heart is too weak to pump blood adequately on its own.
Product code
DSQ Ventricular (Assist) Bypass
Reason for recall
Discovery of serious injuries and deaths associated with the process of changing from a primary System controller to their back-up System controller in patients using the Pocket System controller model.
Root cause tag
design
FDA root cause
Device Design
Action
Thoratec sent an Urgent Medical Device Correction letter dated March 4, 2014 to all affected customers. The letter was sent via e-mail. All distributors outside of the USA and Canada to whom affected product has been distributed have been notified of this field correction notice and have been provided copies of the Customer Letter to forward to their customers. The letter identified the affected product , problem and actions to be taken. Customers were requested to complete and sign the attached Acknowledgement Form and return it to Thoratec via fax (925-847-8571) or e-mail a scanned copy to [email protected]. For questions call 1-800-528-2577.
Quantity in commerce
7436 in total
Distribution
Worldwide distribution: US (nationwide) including states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI; and countries of: Austria, Belgium, Czech Republic, France, Germany, Ireland, Israel, Italy, Lebanon, Lithuania, Poland, Saudi Arabia, Singapore, Slovenia, Sweden, Switzerland, The Netherlands, Turkey, and United Kingdom.
Product codes and lots
Catalog number 105109 (found on side of unit) all serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1230-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1227-2014Class ITerminatedVoluntary: Firm initiated
Z-1228-2014Class ITerminatedVoluntary: Firm initiated
Z-1230-2014Class ITerminatedVoluntary: Firm initiated
Z-229-2014Class ITerminatedVoluntary: Firm initiated