Z-1607-2014Class II

HeartWare HVAD Pump Implant Kit (Heartware Ventricular Assist Device) - HeartWare Battery For use as a...

HeartWare, Inc. / initiated 2014-04-16 / Terminated

HeartWare Inc began a recall of HeartWare HVAD Pump Implant Kit (Heartware Ventricular Assist Device) - HeartWare Battery For use as a bridge-to-cardiac transplantation in patients who are at... on . FDA classified it as Class II and gives the reason as "Patients and caregivers are instructed that if a battery does not provide two hours of support or behaves unusually, that battery should no longer be used and should be replaced". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1607-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
HeartWare, Inc.
Product
HeartWare HVAD Pump Implant Kit (Heartware Ventricular Assist Device) - HeartWare Battery For use as a bridge-to-cardiac transplantation in patients who are at risk of death from refractory end-stage ventricular heart failure.The HeartWare VAS is designed for in-hospital and out-of-hospital settings, including transportation via fix-wing aircraft or helicopter.
Product code
DSQ Ventricular (Assist) Bypass
Reason for recall
Patients and caregivers are instructed that if a battery does not provide two hours of support or behaves unusually, that battery should no longer be used and should be replaced.
FDA root cause
Under Investigation by firm
Action
HeartWare sent an Urgent Medical Device Correction letter dated April 30, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to familiarize themselves with the letter and review their Patient Manual about proper power management. Discuss with their VAD Coordinator any questions they may have about the HeartWare System or the letter. Always keep two (2) power sources connected to your controller; never leave your controller connected to only one (1) power source except while briefly switching to another power source. " Review the recommended practices for power management provided later in this letter. If a battery shows any of the abnormal battery behaviors described, stop using that battery and contact your VAD Coordinator immediately to replace it. Customers were instructed to confirm that they have received and understood the information. Sign and return the enclosed Confirmation form as soon as they can to HeartWare in the prepaid self-addressed envelope or by fax to 1-305-364-2665. For questions customers were instructed to contact their doctor or VAD coordinator. For questions regarding this recall call 305-364-1575.
Quantity in commerce
32,349
Distribution
Worldwide Distribution - USA including AL, AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KY, LA, MA, MD, MI, MN, MO, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia and Internationally to Canada, Dominican Republic, Australia, Belgium, Italy, Germany, New Zeland, South Afica, Japan, Venezuela, Czech Republic, France, Switzerland, Brazil, UK, Saui Arabia, Taiwan, Slovakia, Luxemburg, Hong Kong, Lebano, India, Chile, Oldenburg, Israel, Netherlands, Croatia, Egypt, Turkey, Malaysia, Norway, Greece, Spain, Lithuania, Korea, Singapore, Poland, and Argentina.
Product codes and lots
Catalog# USA:1650 // OUS: 1650, 1650-DE All HeartWare Battery Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1607-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1607-2014Class IITerminatedVoluntary: Firm initiated