The Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationary...
Claymount Assembli began a recall of The Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationary grid. It contains signaling electronics to indicate... on . FDA classified it as Class II and gives the reason as "Claymount Assemblies BV is recalling the Harmony2 model of Harmony Bucky because 1) the devices are incorrectly labeled in that the label does not list the manufacturer address, the date of manufacture, or a statement...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1836-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varex Imaging Nederland B.V.
- Product
- The Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationary grid. It contains signaling electronics to indicate the presence of the grid and detector as well as detector orientation. It may be mounted to a wall stand or incorporated in a radiographic table. It is intended to be used in general radiographic applications as a component of a radiographic system.
- Product code
- IXY Holder, Radiographic Cassette, Wall-Mounted
- Reason for recall
- Claymount Assemblies BV is recalling the Harmony2 model of Harmony Bucky because 1) the devices are incorrectly labeled in that the label does not list the manufacturer address, the date of manufacture, or a statement of compliance with 21 CFR Subchapter J, and 2) the device's insulation values do not comply with IEC 60601-1 in all cases.
- Root cause tag
- labeling and UDI
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- MEDICAL DEVICE RECALL letters dated March 15, 2014 were sent to all direct accounts. The letters includes instructions to: 1) complete and return the Acknowledgement and Receipt Form accompanying the letter; 2) replace the Harmony2 Bucky(s) with new Harmony3 Bucky(s) that the direct account will be receiving in the near future; 3) if the direct account is a distributor and/or installer and it elects to have their customer(s) perform the replacement, the customer(s) should be provided with a copy of the letter, the replacement Harmony3 Bucky(s), and the replacement notification card, or the direct account can provide the names and addresses of their customers in which case Claymount will contact the customers directly; and, 4) once each unit is replaced, the replacement completion notification card should be mailed or faxed to the Claymount USA office and the Harmony2 Bucky(s) should be returned to "Claymount Americas Corp., ATTN: Alice Packard, 2545 Curtiss St., Downers Grove, IL 60515-4059" or faxed to 630-271-9995. For questions customers were instructed to call the following numbers: USA (630) 271-9729 EUR +31 315 659 150
- Quantity in commerce
- 12 devices
- Distribution
- Worldwide Distribution - USA including NC, OH, CO, CA, NJ, IL, OH, PA, FL and Internationally to Canada, China, France, Italy, Netherlands, South Africa, Switzerland, Turkey, and United Kingdom.
- Product codes and lots
- *** US Devices Only *** 1) Part ID: 17360; Serial Numbers: WO186722/E-1 , WO196470/E-1 , WO198534/E-1. 2) Part ID: 17361; Serial Numbers: WO186723/E-1. 3) Part ID: 17362; Serial Numbers: WO186724/E-1 , WO196360/E-1 , WO198545/E-2. 4) Part ID: 18476; Serial Numbers: WO196188/E-1 , WO196188/E-2 , WO196189/E-2. 5) Part ID: 18477; Serial Numbers: WO198547/E-1. 6) Part ID: 18479; Serial Numbers: WO198552/E-1.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1836-2014 | Class II | Terminated | Voluntary: Firm initiated |