'TORAY' Hydrocoat Foley Catheter 12 x10 4.00 mm Fr x mL
Convatec Inc. began a recall of 'TORAY' Hydrocoat Foley Catheter 12 x10 4.00 mm Fr x mL Product Usage: Latex Lubricous Coated Foley Catheters The standard catheter essentially consists of a... on . FDA classified it as Class II and gives the reason as "The lumen of the Medline Sterile Foley Catheter was found to be blocked by excess lubricious coating material". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1240-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ConvaTec Limited
- Product
- 'TORAY' Hydrocoat Foley Catheter 12 x10 4.00 mm Fr x mL Product Usage: Latex Lubricous Coated Foley Catheters The standard catheter essentially consists of a flexible lubricous coated latex tube which incorporates one major drainage lumen and one or two minor lumens constructed in the wall of the tube. One minor lumen is for inflating the retention balloon (available in 2-Way & 3-Way catheters). The second minor lumen (applicable for 3 Way catheters only) is used for bladder irrigation purposes.
- Product code
- KOD Catheter, Urological
- Reason for recall
- The lumen of the Medline Sterile Foley Catheter was found to be blocked by excess lubricious coating material.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Convatec sent Urgent: Voluntary Medical Device Recall letters dated 2/7/2014 via UPS and email. Foreign customers were provided the notice via DHL. The letter identified the affected product, problem and actions to be taken. Customers are asked to perform a count of recalled product in inventory. All consignees should be contacted and be instructed to return product based on their procedures. Customers were instructed to complete and return the enclosed respoonse form.
- Quantity in commerce
- 54,590 catheters
- Distribution
- Worldwide Distribution and in the state of TX
- Product codes and lots
- 540204R001
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1236-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1237-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1238-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1239-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1240-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1241-2014 | Class II | Terminated | Voluntary: Firm initiated |