Sutures, Non-absorbable, Synthetic, Polyethylene, Sterile, Rx only
Teleflex Medical began a recall of Sutures, Non-absorbable, Synthetic, Polyethylene, Sterile, Rx only, Product Usage: Non-absorbable poly(ethylene terephthalate) surgical suture is a... on . FDA classified it as Class II and gives the reason as "The products are being recalled because they did not meet minimum needle attachment strength requirements". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1292-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teleflex Medical
- Product
- Sutures, Non-absorbable, Synthetic, Polyethylene, Sterile, Rx only, Product Usage: Non-absorbable poly(ethylene terephthalate) surgical suture is a multifilament, non-absorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation.
- Product code
- GAT Suture, Nonabsorbable, Synthetic, Polyethylene
- Reason for recall
- The products are being recalled because they did not meet minimum needle attachment strength requirements.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Teleflex sent an Urgent Medical Device Recall Notification letter dated March 11, 2014 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to return all affected product to Teleflex Medical per the instructions provided in the letter. Customers were asked to complete the enclosed Recall Acknowledgment Form and fax it to 1-866-804-9881, Attn: Customer Service. For questions contact your local sales representative or Customer Service at 1-866-246-6990
- Quantity in commerce
- Total 32,271 ea.
- Distribution
- Worldwide Distribution - U.S. Nationwide and the countries of Canada, Germany, and Ireland
- Product codes and lots
- Product Code: 0100019-507, Lot number: 02F1301128; Product Code: 6-511, Lot number: 02M0800561; Product Code: 6-559, Lot number: 02A0800005; Product Code: 69-403, Lot number: 02H0802530; Product Code: 7-5008M4, Lot number: 02C0900466; Product Code: 7-518, Lot number: 02E1302561; Product Code: 7-565, Lot number: 02B1100185; Product Code: 7-655A, Lot numbers: 02C0901963 & 02M0800836; Product Code: 7-740, Lot number: 02L1000536; Product Code: 833-114, Lot numbers: 02A1202112, 02C0903374 & 02D1202794; Product Code: 02F0902697, 02K1100404, 02L1202369 & 02M0802509; Product Code: E13-6351, Lot number: 02F0902446; Product Code: E13-6354, Lot number: 02F0902439; Product Code: E13-6399, Lot number: 02F0902436; Product Code: E6-545, Lot number: 02F0902443; Product Code: E7-4578, ot number: 02D0901672; Product Code: H5300, Lot number: 1450153E13; Product Code: RN6-5106M5, Lot number: 02F1003837; Product Code: RN7-536M5, Lot number: 02A0801205; Product Code: TEV100, Lot number: 02G1101500; Product Code: V-2599, Lot number: 02F0802055; Product Code: X-5424, Lot number: 02B0900765; Product Code: X6-692W, Lot number: 02C0803135; Product Code: X7-655M6A, Lot number: 02A0900806, 02A0902742, 02C0900446 & 02H1003233 and Product Code: XF7-7011, Lot numbers: 02A0901594.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1291-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1292-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1293-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1294-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1295-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1296-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
