Sutures, Non-absorbable, Synthetic, Polypropylene, Sterile, Rx only
Teleflex Medical began a recall of Sutures, Non-absorbable, Synthetic, Polypropylene, Sterile, Rx only, Product Usage: Non-absorbable polypropylene surgical suture is a monofilament... on . FDA classified it as Class II and gives the reason as "The products are being recalled because they did not meet minimum needle attachment strength requirements". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1293-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teleflex Medical
- Product
- Sutures, Non-absorbable, Synthetic, Polypropylene, Sterile, Rx only, Product Usage: Non-absorbable polypropylene surgical suture is a monofilament, Non-absorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation.
- Product code
- GAT Suture, Nonabsorbable, Synthetic, Polyethylene
- Reason for recall
- The products are being recalled because they did not meet minimum needle attachment strength requirements.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Teleflex sent an Urgent Medical Device Recall Notification letter dated March 11, 2014 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to return all affected product to Teleflex Medical per the instructions provided in the letter. Customers were asked to complete the enclosed Recall Acknowledgment Form and fax it to 1-866-804-9881, Attn: Customer Service. For questions contact your local sales representative or Customer Service at 1-866-246-6990
- Quantity in commerce
- Total 32,271 ea.
- Distribution
- Worldwide Distribution - U.S. Nationwide and the countries of Canada, Germany, and Ireland
- Product codes and lots
- Product Code: 833-123, Lot numbers: 02A0902858, 02D0900775, 02G1003092, 02H1100535, 02H1100536, & 02K1102309; Product Code: 833-124, Lot numbers: 02A1200503, 02B1101450 & 02C0903400; Product Code: 02D0900010, 02D0900103, 02D1003285, 02D1003286 & 02F1302321; Product Code: 02G1000045, 02G1300348, 02G1301122, 02H1100494, & 02L0803407; Product Code: 02L1100009, 02M1002289, 02M1101933 & 02F1302322; 833-213, Lot number: 02H1100687; Product Code: D-5007K, Lot numbers: 02B1002310, 02C1103731, 02F1100069 & 02H1102294; Product Code: D-5007M4A, Lot number: 02M0902844; Product Code: D-5007M4K, Lot numbers: 02C1002252, & 02F1100124; Product Code: D-7016M4K, Lot number: 02G1301749; Product Code: D-7070K, Lot numbers: 02A1103450 & 02B1002276: Product Code: D-7070M4K, Lot numbers: 02C1103707, 02F1301100, 02G1100876 & 02G1301739; Product Code: D-7076M1K, Lot number: 02H1103237; Product Code: D-7076M4K, Lot numbers: 02F1101036, 02J1301343, and 02K0900010: Product Code: D-7375K, Lot number: 02A1201015; Product Code: D-793M4K, Lot number: 02L1002488; Product Code: ED-6072, Lot numbers: 02C1002218, 02E1002342 & 02J0902517; Product Code: ED-6276, Lot number: 02F0902457; Product Code: ED-6896, Lot numbers: 02A0902278, 02B0900089, 02B0901762, 02C0900661, 02C1002207, 02D0900634, 02D1100186, 02E0901921, 02E0902608, 02E1301581, 02H1300608, 02J0900501, 02K0901590, 02K0902406, 02L0900676, 02M0901869, & 02G1301755; Product Code: ED-853, Lot numbers: 02G1002594, 02B0902976, 02D0902457 & 02M0901348 and Product Code: EP4049N, Lot number: 02A1003137.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1291-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1292-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1293-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1294-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1295-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1296-2014 | Class II | Terminated | Voluntary: Firm initiated |