510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1317-2014Class II

Citation TMZF HA 132 degrees Neck Angle, V40 Hip Stem - Right Size #3: catalog # 6265-5113 , Left Size #4...

Howmedica Osteonics Corp / initiated 2014-02-03 / Terminated / root cause: labeling and UDI

Stryker Howmedica Osteonics Corp. began a recall of Citation TMZF HA 132 degrees Neck Angle, V40 Hip Stem - Right Size #3: catalog # 6265-5113 , Left Size #4: catalog #6265-5104 , and Right Size #4: catalog... on . FDA classified it as Class II and gives the reason as "Stryker received a report from the field indicating that a citation TMZF HA 132 degrees neck angle V40 Hip Stem - Left Size # 4 was identified in a packaging associated with a citation TMZF HA 132 degrees Neck Angle V40...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1317-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Citation TMZF HA 132 degrees Neck Angle, V40 Hip Stem - Right Size #3: catalog # 6265-5113 , Left Size #4: catalog #6265-5104 , and Right Size #4: catalog #6265-5114; Stryker Howmedica Osteonics Hip Implants.
Product code
MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
Reason for recall
Stryker received a report from the field indicating that a citation TMZF HA 132 degrees neck angle V40 Hip Stem - Left Size # 4 was identified in a packaging associated with a citation TMZF HA 132 degrees Neck Angle V40 Hip Stem - Right Size #3.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Stryker sent an Urgent Medical Device Recall Notification letters and Product Accountability Forms to Hospital Risk Management, Chief of Orthopaedics, and Surgeons via Fed Ex with return receipt on February 26, 2014. The letter identified the affected product, problem and actions to be taken. Customer were instructed to return the affected product to the attention of Regulatory Compliance. Please attach the fluorescent orange PRODUCT REMEDIATION sticker to your return. For questions call 201-972-2100.
Quantity in commerce
48 units
Distribution
USA Nationwide Distribution
Product codes and lots
v40 Hip Stem -Right Size # 3: catalog # 6265-5113 lot code #43317101, v40 Hip Stem -Left Size # 4: catalog #6265-5104 lot code #43316801, lot code 43316901, v40 Hip Stem - Right Size #4: catalog #6265-5114 lot code #43113901

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1317-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1317-2014Class IITerminatedVoluntary: Firm initiated