Proceed Surgical Mesh, Product Codes PCDB1, PCDT1, and PCDJ1. Ethicon Inc., Johnson & Johnson Medical GmbH...
Ethicon, Inc. began a recall of Proceed Surgical Mesh, Product Codes PCDB1, PCDT1, and PCDJ1. Ethicon Inc., Johnson & Johnson Medical GmbH, used for the repair of hernias and other fascial... on . FDA classified it as Class II and gives the reason as "The possibility exists that an incomplete seal on the packaging compromised the sterility of the device and introduced the potential for delamination". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1329-2014
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Ethicon, Inc
- Product
- Proceed Surgical Mesh, Product Codes PCDB1, PCDT1, and PCDJ1. Ethicon Inc., Johnson & Johnson Medical GmbH, used for the repair of hernias and other fascial deficiencies that require the addition of a reinforcing or bridging material to obtain the desired surgical result.
- Product code
- FTM Mesh, Surgical
- Reason for recall
- The possibility exists that an incomplete seal on the packaging compromised the sterility of the device and introduced the potential for delamination.
- Root cause tag
- packaging
- FDA root cause
- Under Investigation by firm
- Action
- Ethicon sent an Urgent Medical Device Recall letter dated February 19, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers are asked to examine their inventory and determine if they have any of the affected product. If any product has been distributed, those hospitals should be notified. The business reply form should be completed and faxed or emailed to Stericycle at 1-888-641-9732 or [email protected]. For clinical or product support contact a customer representative or call the Customer Support Center at 1-877-ETHICON to speak with a registered nurse or contact Stericycle at 1-888-345-0524 and reference event #5798. Stericycle is handling returns.
- Quantity in commerce
- 223 units
- Distribution
- US Distribution including the states of FL, NJ, MO, IL, PA, NC. TX, TN, WA, ID, UT, WV, MS. CA, AL, MA. GA, LA and OK.
- Product codes and lots
- Product Code PCDB1 - Lot number - GGG157 Product Code PCDT1 - Lot number - GGG158 Product Code PCDJ1 - Lot number -GGG159
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1329-2014 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.