Z-1493-2014Class II

Stryker PenAdapt"; Rx Only, Sterile. This device is an accessory to an electrosurgical unit (ESU). The...

Stryker Corp. / initiated 2014-03-26 / Terminated / root cause: manufacturing process

Stryker Instruments Div. of Stryker Corporation began a recall of Stryker PenAdapt"; Rx Only, Sterile. This device is an accessory to an electrosurgical unit (ESU). The PenAdapt is used to remove smoke, particles, and body... on . FDA classified it as Class II and gives the reason as "During packaging verification testing, a failure occurred due to cuts in the packaging potentially causing a breach in the sterile barrier. Additional testing identified microscopic aluminum metal shavings and foreign...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1493-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Stryker PenAdapt"; Rx Only, Sterile. This device is an accessory to an electrosurgical unit (ESU). The PenAdapt is used to remove smoke, particles, and body and casual fluids from the point of surgical activity during surgical procedures that use ESU for cutting and cauterizing.
Product code
FYD Apparatus, Exhaust, Surgical
Reason for recall
During packaging verification testing, a failure occurred due to cuts in the packaging potentially causing a breach in the sterile barrier. Additional testing identified microscopic aluminum metal shavings and foreign particles estimated in size to be between 0.01inches x 0.015 inches to 0.03 inches x 0.04 inches that were introduced during the manufacturing process and were sterilized within the
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
On March 26, 2014, Stryker Instruments sent URGENT MEDICAL DEVICE RECALL NOTIFICATION to their consignees via mail and email. Notification included product description, reason for recall, risk to health, and actions to be taken by the Customer/User. For questions regarding this recall please contact Stryker Instruments: Monday-Friday 8am-5pm (EST) Angela Ragainis/Kelly Jo Whipple 269-389-4354 / 269-389-2921 [email protected] / [email protected] Customers will be asked to sign the Business Reply Form stating they have read the Notification and followed all instructions within. Customers will return affected product to Stericycle for destruction using a return label provided by Stryker or Stericycle. Upon receipt of affected product a credit will be issued to customer account. In addition, facilities are being asked to research their records for patient issues that may have been associated with the use of a recalled device, such as infection and/or inflammation.
Quantity in commerce
35,310 unexpired units
Distribution
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV,and WY, and the countries of Austria, Canada, Sweden, England, UAE, Netherlands, Spain, and Hong Kong.
Product codes and lots
Part Number 0702-045-027

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1493-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1493-2014Class IITerminatedVoluntary: Firm initiated