Z-1445-2014Class II
Alcon LenSx Ophthalmic Laser, for use in cataract surgery for the creation of corneal cuts/incisions...
Alcon Lensx, Inc. / initiated 2014-03-14 / Terminated / root cause: design
Alcon LenSx, Inc. began a recall of Alcon LenSx Ophthalmic Laser, for use in cataract surgery for the creation of corneal cuts/incisions, anterior capsulotomy and laser phacofragmentation on . FDA classified it as Class II and gives the reason as "Alcon LenSx received reports of unexpected downward motion of the gantry". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1445-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alcon Lensx, Inc.
- Product
- Alcon LenSx Ophthalmic Laser, for use in cataract surgery for the creation of corneal cuts/incisions, anterior capsulotomy and laser phacofragmentation.
- Product code
- OOE Ophthalmic Femtosecond Laser
- Reason for recall
- Alcon LenSx received reports of unexpected downward motion of the gantry.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Alcon LenSx Inc advised customers of a Medical Device Correction being initiated for specific serial numbers of the LenSx Laser System for use in cataract surgery for the creation of corneal cuts/incisions, anterior capsulotomy and laser phacofragmentation. Alcon sent customer notification letters via mail during the week of March 17, 2014. The letter titled "Medical Device Correction", advised customers of the correction and provided: the product's description with serial number(s), description of the potential condition, advise on action to be taken by the user, transmission of the notice, and a contact reference person. The letter also contained a section for customers to sign and return the page of the Notice to Alcon.
- Quantity in commerce
- 75
- Distribution
- Worldwide Distribution.
- Product codes and lots
- (US) Serial Numbers: 0311-A007, 0511-A013, 0611-A017, 0611-A018, 0711-A021, 0811-A026, 0811-A027, 0811-A033, 0811-A035, 0911-A039, 0911-A045, 1011-A050, 1011-A062, 1111-A067, 1111-A073, 1111-A077, 1111-A080, 1111-A085, 1211-A086, 1211-A087, 1211-A088, 1211-A090, 0112-A099, 0212-A111, 0212-A116, 0212-A117, 0312-A126, 0312-A130, 0312-A135, 0412-A147, 0412-A152, 0412-A153, 0412-A155, 0512-A163, 0512-A167, 0512-A169. (International) Serial Numbers: 0611-A020, 0711-A022, 0711-A025, 0811-A030, 0811-A031, 0811-A032, 0811-A037, 0911-A038, 0911-A040, 0911-A041, 0911-A042, 1011-A053, 1011-A054, 1011-A055, 1011-A060, 1111-A069, 1111-A071, 1111-A072, 1111-A074, 1111-A075, 1111-A081, 1111-A082, 1211-A091, 1211-A094, 0112-A106, 0112-A107, 0112-A108, 0112-A109, 0212-A118, 0312-A121, 0312-A122, 0312-A123, 0312-A127, 0312-A132, 0512-A158, 0512-A160, 0512-A162, 0512-A164, 0512-A166.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1445-2014 | Class II | Terminated | Voluntary: Firm initiated |