Z-1464-2014Class II
Ingenuity Core 128, Computed Tomography X-ray systems, Philips Medical Systems (Cleveland), Inc
Philips North America LLC / initiated 2014-03-06 / Terminated / root cause: software
Philips Medical Systems (Cleveland) Inc began a recall of Ingenuity Core 128, Computed Tomography X-ray systems, Philips Medical Systems (Cleveland), Inc on . FDA classified it as Class II and gives the reason as "A customer reported that after upgrading to software version 3.5.5 from 3.5.4, reconstructions would intermittently not complete". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1464-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Ingenuity Core 128, Computed Tomography X-ray systems, Philips Medical Systems (Cleveland), Inc.
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- A customer reported that after upgrading to software version 3.5.5 from 3.5.4, reconstructions would intermittently not complete.
- Root cause tag
- software
- FDA root cause
- Software in the Use Environment
- Action
- .The firm, Phillips, sent an " URGENT Medical Device Correction" letter dated 2013 March 06, to its costumers . The letter disc ribes the product, problem and actions to be taken. The costumers were instructed to Verify that all planned image reconstructions are completed, and Perform off-line reconstructions if expected images are missing. The firm will release A Field Change Order (FCO) to correct the issue with software version release 3.5.7.and A Philips Field Service Engineer will contact you to schedule the software installation at your site. If you need any further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, Option 5: Enter Site ID or follow the prompts).
- Quantity in commerce
- 116 units
- Distribution
- Worldwide Distribution:US ( Nationwide) Including stages of MI, OR, PA, TX & WV; and Countries of: Australia, Bangladesh, Belgium, Brazil, Bulgaria, China, France, Germany, India, Indonesia, Ireland, Japan, Nicaragua, Oman, Palestine, Russia, South Korea, Spain, Switzerland, Thailand, Turkey & United Kingdom.
- Product codes and lots
- System code: 728323; Serial #: 320002, 320003, 320004, 320005, 320006, 320007, 320008, 320009, 320010, 320011, 320013, 320014, 320015, 320017, 320018, 320019, 320020, 320022, 320023, 320024, 320025, 320026, 320027, 320028, 320029, 320031, 320032, 320034, 320035, 320036, 320037, 320038, 320039, 320040, 320041, 320043, 320044, 320045, 320046, 320047, 320048, 320049, 320050, 320051, 320052, 320053, 320054, 320055, 320056, 320057, 320059, 320060, 320062, 320063, 320064, 320065, 320066, 320067, 320068, 320069, 320070, 320071, 320072, 320074, 320075, 320078, 320079, 320080, 320081, 320082, 320083, 320084, 320085, 320086, 320087, 320088, 320089, 320090, 320092, 320093, 320095, 320096, 320097, 320098, 320099, 320101, 320102, 320103, 320104, 320105,320106, 320107, 320109, 320110, 320111, 320112, 320114, 320115, 320116, 320117, 320118, 320119, 320120, 320121, 320122, 336013 & 336015.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1464-2014 | Class II | Terminated | Voluntary: Firm initiated |