Z-1464-2014Class II

Ingenuity Core 128, Computed Tomography X-ray systems, Philips Medical Systems (Cleveland), Inc

Philips North America LLC / initiated 2014-03-06 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of Ingenuity Core 128, Computed Tomography X-ray systems, Philips Medical Systems (Cleveland), Inc on . FDA classified it as Class II and gives the reason as "A customer reported that after upgrading to software version 3.5.5 from 3.5.4, reconstructions would intermittently not complete". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1464-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Ingenuity Core 128, Computed Tomography X-ray systems, Philips Medical Systems (Cleveland), Inc.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
A customer reported that after upgrading to software version 3.5.5 from 3.5.4, reconstructions would intermittently not complete.
Root cause tag
software
FDA root cause
Software in the Use Environment
Action
.The firm, Phillips, sent an " URGENT Medical Device Correction" letter dated 2013 March 06, to its costumers . The letter disc ribes the product, problem and actions to be taken. The costumers were instructed to Verify that all planned image reconstructions are completed, and Perform off-line reconstructions if expected images are missing. The firm will release A Field Change Order (FCO) to correct the issue with software version release 3.5.7.and A Philips Field Service Engineer will contact you to schedule the software installation at your site. If you need any further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, Option 5: Enter Site ID or follow the prompts).
Quantity in commerce
116 units
Distribution
Worldwide Distribution:US ( Nationwide) Including stages of MI, OR, PA, TX & WV; and Countries of: Australia, Bangladesh, Belgium, Brazil, Bulgaria, China, France, Germany, India, Indonesia, Ireland, Japan, Nicaragua, Oman, Palestine, Russia, South Korea, Spain, Switzerland, Thailand, Turkey & United Kingdom.
Product codes and lots
System code: 728323; Serial #: 320002, 320003, 320004, 320005, 320006, 320007, 320008, 320009, 320010, 320011, 320013, 320014, 320015, 320017, 320018, 320019, 320020, 320022, 320023, 320024, 320025, 320026, 320027, 320028, 320029, 320031, 320032, 320034, 320035, 320036, 320037, 320038, 320039, 320040, 320041, 320043, 320044, 320045, 320046, 320047, 320048, 320049, 320050, 320051, 320052, 320053, 320054, 320055, 320056, 320057, 320059, 320060, 320062, 320063, 320064, 320065, 320066, 320067, 320068, 320069, 320070, 320071, 320072, 320074, 320075, 320078, 320079, 320080, 320081, 320082, 320083, 320084, 320085, 320086, 320087, 320088, 320089, 320090, 320092, 320093, 320095, 320096, 320097, 320098, 320099, 320101, 320102, 320103, 320104, 320105,320106, 320107, 320109, 320110, 320111, 320112, 320114, 320115, 320116, 320117, 320118, 320119, 320120, 320121, 320122, 336013 & 336015.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1464-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1464-2014Class IITerminatedVoluntary: Firm initiated