Z-1460-2014Class II

AXIOM Vertix MD Trauma systems radiographic X-ray

Siemens Medical Solutions USA, Inc. / initiated 2014-03-05 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of AXIOM Vertix MD Trauma systems radiographic X-ray on . FDA classified it as Class II and gives the reason as "There is a potential issue and possible hazard to patients when using the AXIOM Vertix MD Trauma systems. In rare cases, steel ropes inside the lift column of the system can be defective without triggering the safety...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1460-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
AXIOM Vertix MD Trauma systems radiographic X-ray
Product code
IZZ Table, Radiographic, Non-Tilting, Powered
Reason for recall
There is a potential issue and possible hazard to patients when using the AXIOM Vertix MD Trauma systems. In rare cases, steel ropes inside the lift column of the system can be defective without triggering the safety lock, which can result in the U-arm dropping down unexpectedly during movement in vertical direction, potentially causing serious injury.
Root cause tag
design
FDA root cause
Component design/selection
Action
Siemens sent a Safety Advisory Notice dated March 5, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were advised as a first check it is strongly recommended for the users to check whether metallic dust or rubbed off parts of metal are visible underneath the lifting column or around the system. If this is the case, it is strongly recommended to immediately stop using the Vertix MD Trauma system and call the local Siemens service. To avoid any risk until the implementation of the modification mentioned below, it is furthermore strongly recommended to perform up/down movements of the lifting column not directly above the patient, but complete the vertical movement beside the patient and then move the system horizontally above the patient. We appreciate your understanding and cooperation with this Safety Advisory Notice and ask you to immediately instruct your personnel accordingly. Please ensure that this Safety Advisory Notice is placed in the system's instructions for use until the update has been installed. If you have sold or otherwise disposed of this equipment and it is no under your control, we kindly ask that you forward this Safety Advisory Notice to the new user of the equipment. Please also inform us about the new owner of the equipment. We apologize for any inconvenience this may cause. Further questions please call (610) 219-6300.
Quantity in commerce
2
Distribution
US Distribution including MO and OH.
Product codes and lots
AXIOM Vertix MD Trauma systems (material no. 08908290) with serial numbers 1022 through 1058.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1460-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1460-2014Class IITerminatedVoluntary: Firm initiated