Z-1442-2014Class I

SpaceLabs Arkon Anesthesia Delivery System with Version 2.0 Software installed. This system includes a...

GE Healthcare / initiated 2014-03-10 / Terminated / root cause: software

Del Mar Reynolds Medical, Ltd. began a recall of SpaceLabs Arkon Anesthesia Delivery System with Version 2.0 Software installed. This system includes a ventilator on . FDA classified it as Class I and gives the reason as "The System may go into a controlled failed state if a series of parameters on the user interface are activated in a specific sequence". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1442-2014
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
SpaceLabs Arkon Anesthesia Delivery System with Version 2.0 Software installed. This system includes a ventilator.
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
The System may go into a controlled failed state if a series of parameters on the user interface are activated in a specific sequence.
Root cause tag
software
FDA root cause
Software in the Use Environment
Action
Spacelabs Healthcare sent an Urgent Medical Device Correction letter, dated 11 March 2014, to consignees. The letter identified the product, the problem, and the actions to be taken by the customer. The letter stated that Spacelabs Field Service personnel will be contacting consignees to schedule a convenient time to install, at no cost, a software upgrade. Consignees with questions about this corrective action program, please contact Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical Support.
Quantity in commerce
16 units distributed in the US
Distribution
USA Nationwide Distribution in the states of North Carolina and South Carolina
Product codes and lots
SERIAL NUMBERS OF UNITS DISTRIBUTED IN THE US:  ARKN-000011, ARKN-000016, ARKN-000017, ARKN-000019, ARKN-000020, ARKN-000021, ARKN-000022, ARKN-000023, ARKN-000024, ARKN-000025, ARKN-000026, ARKN-000027, ARKN-000028, ARKN-000029, ARKN-000030, and ARKN-000031.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1442-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1442-2014Class ITerminatedVoluntary: Firm initiated