Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1491-2014Class III

The CEP 8 SpectrumGreen (SG) ASR Probe Kit, 20ul (list 06J37-018) is an Analyte Specific Reagent (ASR)...

Abbott Molecular, Inc. / initiated 2013-03-28 / Terminated / root cause: packaging

Abbott Molecular began a recall of The CEP 8 SpectrumGreen (SG) ASR Probe Kit, 20ul (list 06J37-018) is an Analyte Specific Reagent (ASR) containing a single vial of CEP 8 spectrum green probe... on . FDA classified it as Class III and gives the reason as "20 kits of CEP 8 SG ASR Probe (list 06J37-018, lot 440023), contained Vysis CEP 4 (Alpha Satellite) SG Probe, 20u, (part 32-112004 lot 438814), instead of the expected CEP 8 probe". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1491-2014
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Molecular, Inc.
Product
The CEP 8 SpectrumGreen (SG) ASR Probe Kit, 20ul (list 06J37-018) is an Analyte Specific Reagent (ASR) containing a single vial of CEP 8 spectrum green probe and a Certificate of Analysis (COA). Chromosome enumeration probes (CEP) are chromosome-specific FISH probes that hybridize to highly repetitive human satellite DNA sequences, usually located near centromeres. CEP signals enable the identification and enumeration of human chromosomes in interphase and metaphase cells from fresh and archived samples. The label for CEP 8 SG ASR Probe Kit indicates it is an Analyte Specific Reagent, and analytical and performance characteristics are not established.
Product code
MVU Reagents, Specific, Analyte
Reason for recall
20 kits of CEP 8 SG ASR Probe (list 06J37-018, lot 440023), contained Vysis CEP 4 (Alpha Satellite) SG Probe, 20u, (part 32-112004 lot 438814), instead of the expected CEP 8 probe.
Root cause tag
packaging
FDA root cause
Packaging change control
Action
Customers were sent Field Correction Notices dated 3/28/13 instructing them to inspect their CEP 8 SpectrumGreen Probe kits for the presence of CEP4 probe vials. If any CEP4 vials are discovered, customers should contact their local Abbott Molecular Technical Support Represtative for coordination of reagent replacement with the correct CEP8 vial. If customers have any specific questions they can contact Abbott Molecular at 1-800-553-7042, Option 2.
Quantity in commerce
20 kits
Distribution
Nationwide (CA, CO, CT, FL, HI, IL, IN, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI); Canada
Product codes and lots
List Number: 06J37-018; Lot Number: 440023

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1491-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1491-2014Class IIITerminatedVoluntary: Firm initiated