510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1452-2014Class II

Reusable latex breathing bags with multiple part numbers and sold as part of test lung part number and...

Draeger, Inc. / initiated 2014-03-25 / Terminated / root cause: labeling and UDI

Draeger Medical, Inc. began a recall of Reusable latex breathing bags with multiple part numbers and sold as part of test lung part number and reusable anaesthesia sets. Product Usage: The breathing... on . FDA classified it as Class II and gives the reason as "Reusable latex breathing bags with various part numbers and sold as part of test lung and reusable anaesthesia sets were recalled when Draeger became aware of an incident where a user suffered an acute allergic reaction...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1452-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Draeger, Inc.
Product
Reusable latex breathing bags with multiple part numbers and sold as part of test lung part number and reusable anaesthesia sets. Product Usage: The breathing bag is a component of the anesthesia set (breathing circuit) that is intended to is the conduit for administering medical gases to a patient during ventilation. It provides both an inspiratory and expiratory route. 2. The test lung is an apparatus that contains a breathing bag and and connector. It is connected to the output of a ventilator and is used as part of the device/system check out.
Product code
BZO Set, Tubing And Support, Ventilator (W Harness)
Reason for recall
Reusable latex breathing bags with various part numbers and sold as part of test lung and reusable anaesthesia sets were recalled when Draeger became aware of an incident where a user suffered an acute allergic reaction after using a Drger latex test lung. The user did not suffer permanent injury. The breathing bag that belongs to the test lung and other latex breathing bags are labelled NR. A
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
A recall notification letter, dated March 2014, was sent to wholesale distributors and end users.
Quantity in commerce
433
Distribution
USA Nationwide Distribution in the states of: MI, SC, CO, OH, NE, WI, NC, SD, AZ, TX, MA, WA, PA, KY, MN, GA, OK, NY, IN, VA, UT, MT, ME,FL, CA, MD, MO, IA, TN, MD, WV, IL,NJ, CT, and AR.
Product codes and lots
Reusable latex breathing bags with part numbers 2165686, 2165953, 2165694, as part of test lung part number 8403201, and reusable anaesthesia sets part numbers M33681 and M27542.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1452-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1452-2014Class IITerminatedVoluntary: Firm initiated