Z-1633-2014Class II

Arrowgard Blue and Arrowgard Blue Plus Pressure Injectable Central Venous Catheter, a multiple-lumen catheter...

Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2013-07-17 / Terminated / root cause: labeling and UDI

Arrow International Inc began a recall of Arrowgard Blue and Arrowgard Blue Plus Pressure Injectable Central Venous Catheter, a multiple-lumen catheter that permits venous access to central... on . FDA classified it as Class II and gives the reason as "The labeling states that the kit contains Introducer Needle: Echogenic 18 Ga. x 2.5 (6.35cm) XTW and 5mL ArroW Raulerson Spring-Wire Introduction Syringe; however, the kit contains, and should state, Introducer Needle...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1633-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arrow International LLC (subsidiary of Teleflex Incorporated)
Product
Arrowgard Blue and Arrowgard Blue Plus Pressure Injectable Central Venous Catheter, a multiple-lumen catheter that permits venous access to central circulation. Catalog number ASK-45703-PHF.
Product code
FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Reason for recall
The labeling states that the kit contains Introducer Needle: Echogenic 18 Ga. x 2.5 (6.35cm) XTW and 5mL ArroW Raulerson Spring-Wire Introduction Syringe; however, the kit contains, and should state, Introducer Needle: 18 Ga. x 2.5 (6.35cm) XTW and 5mL ArroW Raulerson Spring-Wire Introduction Syringe. As a result of this labeling inconsistency, if the clinician were to place the needle under x-r
Root cause tag
labeling and UDI
FDA root cause
Labeling False and Misleading
Action
Arrow sent an Urgent Medical Device Recall Notification letter dated July 17, 2013, was sent to the customer requesting they discontinue use and quarantine any remaining affected product for return to Arrow. Our records indicate that you have received products that are subject to this voluntary recall. We are now asking that you take the following actions: 1. If you have affected stock, immediately discontinue use and quarantine any products with the product and lot numbers listed above. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-800-343-2935, Attn: Customer Service. This will allow us to document the amount of product you have on hand for return. Your sales representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions and assistance for the return of product to Arrow International. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-800-343-2935, Attn: Customer Service. This will allow us to document your receipt of this letter.
Quantity in commerce
4285
Distribution
US Distribution only to MI.
Product codes and lots
MF2016829 RM2072044 RM2034076 RM2095792 RM2056629 RM3015571 RM2069733

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1633-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1633-2014Class IITerminatedVoluntary: Firm initiated