Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1330-2014Class II

Spirit MB Bracket, UR Central +14T +5A.018, Part Number 494-0110, Lot Number 081351013

Ormco/Sybronendo / initiated 2014-03-25 / Terminated

Ormco/Sybronendo began a recall of Spirit MB Bracket, UR Central +14T +5A.018, Part Number 494-0110, Lot Number 081351013. Product Usage: An orthodontic bracket. Spirit MB Brackets are a fixed... on . FDA classified it as Class II and gives the reason as "Ormco Corporation initiated this recall of Spirit MB Brackets, Part Number 494-0110, Lot Number 081351013, because the ID dot, which designates the bracket as distal/gingival, is misplaced. The misidentification may...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1330-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ormco/Sybronendo
Product
Spirit MB Bracket, UR Central +14T +5A.018, Part Number 494-0110, Lot Number 081351013. Product Usage: An orthodontic bracket. Spirit MB Brackets are a fixed attachment on a tooth which holds an archwire during orthodontic treatment.
Product code
DYW Bracket, Plastic, Orthodontic
Reason for recall
Ormco Corporation initiated this recall of Spirit MB Brackets, Part Number 494-0110, Lot Number 081351013, because the ID dot, which designates the bracket as distal/gingival, is misplaced. The misidentification may lead an orthodontist to place the bracket on the wrong tooth, leading to movement 10 degrees in the wrong direction.
FDA root cause
Use error
Action
Ormco Corporation initiated this recall by sending recall letters to customers on March 25, 2014 via 1st class mail. The letter identified the affected product, problem and actions to be taken. The consignees are instructed to complete the "Return Form" and return any affected product in their inventory. The letter was written in Japanese language and accompanied by a "Return Form".
Quantity in commerce
150 units
Distribution
Internationally Distributed in the country of Japan.
Product codes and lots
Part Number 494-0110, Lot Number 081351013.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1330-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1330-2014Class IITerminatedVoluntary: Firm initiated