510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1649-2014Class II

Esmark Elastic Bandage (Sterile), 4" x 9'. Item DYNJ05116A, Individually wrapped, 20 bandages per case...

MEDLINE INDUSTRIES, LP - Northfield / initiated 2014-03-20 / Terminated / root cause: manufacturing process

Medline Industries Inc began a recall of Esmark Elastic Bandage (Sterile), 4" x 9'. Item DYNJ05116A, Individually wrapped, 20 bandages per case. Packaged by Medline Industries Inc. This product is... on . FDA classified it as Class II and gives the reason as "This lot did not go through the correct sterilization procedures. This product may potentially be non-sterile". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1649-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Esmark Elastic Bandage (Sterile), 4" x 9'. Item DYNJ05116A, Individually wrapped, 20 bandages per case. Packaged by Medline Industries Inc. This product is used as an elastic bandage to support and compress a part of a patient's body. It is also used as a tourniquet to restrict blood flow to a part of a patient's body.
Product code
FQM Bandage, Elastic
Reason for recall
This lot did not go through the correct sterilization procedures. This product may potentially be non-sterile.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Medline Industries issued an Immediate Action Required letter dated March 20, 2014 to all affected customers. The letters included instructions to: 1) immediately check inventories for the recalled products and quarantine the recalled products; 2) complete and return the enclosed destruction form listing the quantity of destroyed product (credit will only be issued if the completed form is received); and 3) If the customer is a distributor, promptly notify the distributor's customers that may have received the recalled products about this recall. Direct accounts and their customers that have any questions can contact Medline Industries at 866-359-1704.
Quantity in commerce
2,860 bandages
Distribution
Nationwide Distribution including AZ, CA, IA, IL, IN, KY, LA, MI, MS, NJ, NY, OH, PA, TX, and WA.
Product codes and lots
Product Number: DYNJ05116A .  Lot Number: 13LA1009.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1649-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1649-2014Class IITerminatedVoluntary: Firm initiated