Z-1495-2014Class II

Philips MDC PACS R2.X / IntelliSpace PACS DCX R3.1 software, software application that is used for receiving...

Philips North America LLC / initiated 2014-04-02 / Terminated / root cause: software

Philips Medical Systems, Inc. began a recall of Philips MDC PACS R2.X / IntelliSpace PACS DCX R3.1 software, software application that is used for receiving, managing, archiving, distributing and recording... on . FDA classified it as Class II and gives the reason as "Philips MDC PACS Release R2.x and Philips IntelliSpace PACS DCX R3.x, may have incorrect density measurement on Enhanced CT/MR examinations". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1495-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips MDC PACS R2.X / IntelliSpace PACS DCX R3.1 software, software application that is used for receiving, managing, archiving, distributing and recording medical images onto portable digital media (including but not limited to Compact Disk and DVD). The MDC PACS/IntelliSpace PACS DCX is a software application that is used for receiving, managing, archiving, distributing and recording medical images onto portable digital media (including but not limited to Compact Disk and DVD). The MDC PACS / IntelliSpace PACS DCX receives digital images and data from various sources (including but not limited to CT, MR, US, NM, XA, RF, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources). Typical users of the MDC PACS /IntelliSpace PACS DCX and DIAGNET are trained medical professionals, including but not limited to radiologists, clinicians. technologist and others
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Philips MDC PACS Release R2.x and Philips IntelliSpace PACS DCX R3.x, may have incorrect density measurement on Enhanced CT/MR examinations
Root cause tag
software
FDA root cause
Software design
Action
Philips Healthcare sent a Urgent Medical Device Correction Notification/Field Safety Notice dated March 2014, to inform customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. Philips will replace the software free of charge. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice: Until you receive a software upgrade from Philips, the user must configure the MRI and CT devices to send regular CT or MR DICOM image objects (not enhanced) to PACS.. This issue does not affect images that are not enhanced image objects. Further questions please call (978) 687-1501.
Quantity in commerce
373 units
Distribution
Worldwide Distribution - US Distribution including Puerto Rico and the states of CA, FL, IL, NJ, OK and WA., and the countries of : Argentina Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Bosnia, Brazil, Bulgaria, Chile, China, Egypt, Ethiopia, France, Germany, Ghana, Greece, Guinea, India, Iraq, Israel, Italy, Latvia, Libya, Lithuania, Malawi, Malta, Mexico, Mozambique, Nepal, New Zealand, Papua, New Guinea, Paraguay, Philippines, Poland, Portugal, Romania, Russia, Slovakia, Slovenia, Solomon Islands, South Africa, Spain, Thailand, Turkey, United Arab Emirates and United Kingdom.
Product codes and lots
Software Versions:  MDC PACS R2.2; R2.3  IntelliSpace PACS DCX R3.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1495-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1495-2014Class IITerminatedVoluntary: Firm initiated