510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1503-2014Class II

Digital Diagnost, stationary radiographic system; Models: 712020/22/25/26/50/52/54/55/57/60/62/80/81/82/85

Philips North America LLC / initiated 2014-03-17 / Terminated / root cause: manufacturing process

Philips Medical Systems, Inc. began a recall of Digital Diagnost, stationary radiographic system; Models: 712020/22/25/26/50/52/54/55/57/60/62/80/81/82/85 on . FDA classified it as Class II and gives the reason as "The footboard of Patient support for stitching can be fixed in upper position. When the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1503-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Digital Diagnost, stationary radiographic system; Models: 712020/22/25/26/50/52/54/55/57/60/62/80/81/82/85
Product code
MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
Reason for recall
The footboard of Patient support for stitching can be fixed in upper position. When the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On March 17, 2014, consignees were notified of the recall via letter. The letter explains what the issue is, actions firms can take to avoid or minimize the occurrence of the issue, and the actions planned by Philips to correct the issue. For further information or support concerning this issue, please contact your local Philips Representative. 1-800-722-9377
Quantity in commerce
985 devices
Distribution
Nationwide: US: AL, AR, AZ, CA, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, WY / Worldwide including Canada & Mexico
Product codes and lots
Units distributed from November 2006 through March 2014.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1503-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1503-2014Class IITerminatedVoluntary: Firm initiated