Z-1530-2014Class II
Alfa Wassermann Diagnostic Technologies, LLC. ACE AST Reagent
Alfa Wassermann / initiated 2014-03-05 / Terminated / root cause: manufacturing process
Alfa Wassermann, Inc. began a recall of Alfa Wassermann Diagnostic Technologies, LLC. ACE AST Reagent. Product Usage: For quantitative determination of AST in serum and lithium heparin plasma using... on . FDA classified it as Class II and gives the reason as "Extremely low absorbance readings". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1530-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alfa Wassermann
- Product
- Alfa Wassermann Diagnostic Technologies, LLC. ACE AST Reagent. Product Usage: For quantitative determination of AST in serum and lithium heparin plasma using ACE, ACE Alera and ACE Axcel Clinical Chemistry Systems. This test is intended for use in clinical laboratories and physician office laboratories.
- Product code
- CIT Nadh Oxidation/Nad Reduction, Ast/Sgot
- Reason for recall
- Extremely low absorbance readings.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Alfa Wassermann Diagnostic Technologies, LLC issued an "Important: Advisory Notice/"Product Response Form" dated March 5, 2014 and sent on March 6, 2014 to their customers for ACE ALT Reagent Product reorder number SA1046 Lot F3604. They issued a follow-up "Important: Advisory Notice/"Product Response Form" dated March 21, 2014 and sent on March 26, 2014 to their customers for ACE AST Reagent Product reorder numbers SA1047, Lot F3609, and RX1047, Lot F3610. The notification identified the product(s), problem, and actions to be taken by the customers. Customers were instructed to destroy any affected product and complete and submit the enclosed Product Response Form. For questions or concerns, contact the Alfa Wassermann Diagnostic Technologies, LLC Customer Solution Center, Toll Free: (866) 419-ALFA (2532).
- Quantity in commerce
- SA1047 lot F3609: 146; RX1047 lot F3610: 53
- Distribution
- US Nationwide Distribution
- Product codes and lots
- ACE¿ AST Reagent Product reorder numbers SA1047 (Lot F3609) and RX1047 (Lot F3610)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1529-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1530-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.