Z-1516-2014Class II

The Acumed Tension Band Pin is a fixation pin which is .0625" in diameter and is available in lengths between...

Acumed LLC / initiated 2014-03-26 / Terminated / root cause: manufacturing process

Acumed LLC began a recall of The Acumed Tension Band Pin is a fixation pin which is .0625" in diameter and is available in lengths between 35mm and 90mm. It is to be used in conjunction... on . FDA classified it as Class II and gives the reason as "Certain lots of Acumed Tension Band Pins are being recalled because the implantable product was manufactured with a grade of stainless steel that is not within specifications". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1516-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Acumed LLC
Product
The Acumed Tension Band Pin is a fixation pin which is .0625" in diameter and is available in lengths between 35mm and 90mm. It is to be used in conjunction with commonly used orthopedic wire. 50.0 mm Tension Band Pin; Part Number 30-0097 70.0 mm Tension Band Pin; Part Number 30-0098 90.0 mm Tension Band Pin; Part Number 30-0099 50.0 mm Tension Band Kit; Part Number TB-1550K-S 70.0 mm Tension Band Kit; Part Number TB-1570K-S 90.0 mm Tension Band Kit; Part Number TB-1590K-S Product Usage: Acumed Tension Band Pin is intended to be used with orthopedic wire to address malleolar, patella, and olecranon fractures in tension band wiring procedures. This device is not intended for usage in the spine.
Product code
HRS Plate, Fixation, Bone
Reason for recall
Certain lots of Acumed Tension Band Pins are being recalled because the implantable product was manufactured with a grade of stainless steel that is not within specifications.
Root cause tag
manufacturing process
FDA root cause
Component change control
Action
Acumed sent an Urgent Medical Device Recall letter dated March 26, 2014 to all affected customers. The letter identified the affected products, problem and actions to be taken. Customers are instructed to immediately stop using and/or distributing all lots identified in the letter. Return affected products as instructed in the letter. For questions contact Acumed Business Service via email at [email protected] or via phone at 877-627-9957.
Quantity in commerce
2544 unit in the US and 3584 units outside the US
Distribution
US Nationwide Distribution - in the states of CA, CO, FL, GA, LA, MI, MN, MO, OH, OR, PA, PR, TX, UT, VA, WA and WY. Product was distributed outside the US to: Australia, China, Figi, France, Great Britain, Italy, Japan, Korea, Maylasia, Sweden, South Africa, Spain.
Product codes and lots
50.0 mm Tension Band Pin: Lot 243639 70.0 mm Tension Band Pin: Lots 241127, 242707, 243325, 243948. 90.0 mm Tension Band Pin: Lots 241128. 242708, 243327, 243949.  50.0 mm Tension Band Kit: Lots 216769, 217002, 217941, 219145, 219402, 222732, 223007, 224312, 229889, 232575, 235490, 237772, 244800, 246204, 247056.  70.0 mm Tension Band Kit: Lots 210290, 211349, 214574, 215405, 217004, 218273, 219230, 220978, 221503, 223008, 223477, 224066, 225554, 227193, 229394, 230287, 230952, 231971, 233742, 235502, 237773, 240512, 241185, 243617, 243945, 244801, 246777, 246779, 246938, 247108, 247109, 247493.  90.0 mm Tension Band Kit: Lots 215053, 217005, 218600, 218700, 221465, 222731, 223009, 223756, 225555, 228951, 231573, 236969, 237774, 243620, 244802, 246939, 246990, 247381.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1516-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1515-2014Class IITerminatedVoluntary: Firm initiated
Z-1516-2014Class IITerminatedVoluntary: Firm initiated