510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1652-2014Class II

Siemens Linear Accelerator (LINAC) models

Siemens Medical Solutions USA, Inc. / initiated 2014-04-01 / Terminated / root cause: labeling and UDI

Siemens Medical Solutions USA, Inc began a recall of Siemens Linear Accelerator (LINAC) models. Product Usage: deliver X-ray radiation for therapeutic treatment of cancer on . FDA classified it as Class II and gives the reason as "Siemens Radiation Oncology became aware that customers may be using the Siemens Linear Accelerator in combination with stereotactic accessories which have not been validated as being compatible with Siemens LINAC models". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1652-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Linear Accelerator (LINAC) models. Product Usage: deliver X-ray radiation for therapeutic treatment of cancer
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Siemens Radiation Oncology became aware that customers may be using the Siemens Linear Accelerator in combination with stereotactic accessories which have not been validated as being compatible with Siemens LINAC models.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Siemen sent an Urgent Medical Device Correction letter to all affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local Elekta representative.
Quantity in commerce
9
Distribution
US Nationwide Distribution in the states of AL, ME, PA, IL, and CA.
Product codes and lots
material numbers and serial numbers: 1940035 2962 8139789 5833 8139789 5899 9401654 2113 1940035 2628 4504200 3066 4504200 3361 1940035 3873 1940035 2855

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1652-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1652-2014Class IITerminatedVoluntary: Firm initiated