Z-1487-2014Class II

iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations of...

Boston Scientific Corporation / initiated 2014-04-03 / Terminated / root cause: manufacturing process

Boston Scientific Corporation began a recall of iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations of intravascular pathology on . FDA classified it as Class II and gives the reason as "Specific iLab Ultrasound Imaging Systems using newest 3.0 (Polaris 1.0) software change the pullback speed from the default setting of 0.5 mm/sec to 1.0 mm/sec, the Longview distance measurement may display an...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1487-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations of intravascular pathology.
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
Specific iLab Ultrasound Imaging Systems using newest 3.0 (Polaris 1.0) software change the pullback speed from the default setting of 0.5 mm/sec to 1.0 mm/sec, the Longview distance measurement may display an inaccurate value that is double the actual distance.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Boston Scientific sent an Urgent Medical Device Field Correction letter to all affected customers via hand delivery by BSC sales representatives. The letter identifed the affected product, problem and actions to be taken. Customers were asked to immediately post the Field Correction information in a visible location near the system(s) to ensure it is easily accessible to all users of the device. For question contact your local Sales Representative.
Quantity in commerce
9 units
Distribution
USA Nationwide Distribution in the states of NY, NC, DC, FL, PA
Product codes and lots
Model Numbers: H749ILAB120CART0 H749ILAB120INS0 H749ILAB120N2710 Serial numbers: 4219, 4900, 5701, 5827 7516, 7517, 7607 8407, 8434. the different UPNs represent minor level differences in configuration relative to mobility and no difference in any essential components or software.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1487-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1487-2014Class IITerminatedVoluntary: Firm initiated