Z-1539-2014Class I

Infant Nasal CPAP (continuous positive airway pressure) Prong, Nasal. CPAP Prongs (10 pack): Model No...

Fisher & Paykel Healthcare, Inc. / initiated 2014-04-15 / Terminated / root cause: design

Fisher & Paykel Healthcare, Ltd. began a recall of Infant Nasal CPAP (continuous positive airway pressure) Prong, Nasal. CPAP Prongs (10 pack): Model No. BC3020-10, BC3520-10, BC4030-10, BC4540-10, BC5040-10... on . FDA classified it as Class I and gives the reason as "The Infant Nasal CPAP Prong has the potential to detach from the Nasal Tubing and therapy is likely to be interrupted". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1539-2014
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fisher & Paykel Healthcare, Inc.
Product
Infant Nasal CPAP (continuous positive airway pressure) Prong, Nasal. CPAP Prongs (10 pack): Model No. BC3020-10, BC3520-10, BC4030-10, BC4540-10, BC5040-10, BC5050-10, BC5550-10, BC5560-10, BC6060-10, BC6070-10, BC6570-10. Bubble CPAP Starter Kits: Model No. BC461-SK, BC471-SK, BC490-SK, BC491-SK, BC492-SK. The Nasal CPAP Prongs connect to the FPH FlexiTrunk Patient Interface.
Product code
BZO Set, Tubing And Support, Ventilator (W Harness)
Reason for recall
The Infant Nasal CPAP Prong has the potential to detach from the Nasal Tubing and therapy is likely to be interrupted.
Root cause tag
design
FDA root cause
Device Design
Action
Fisher & Paykel Healthcare sent an Urgent Medical Device Recall letter on April 11, 2014, to all affected customers. The letter informed them of the problems identified and the actions to be taken. Customers were instructed to complete Section A 'Inspection of Stock' on the Urgent Medical Device Recall Response Form and return it to their FPH Representative. Destroy the Affected Prongs by cutting them in half. Customers were also instructed to contact their FRH Representative for replacement product. For questions regarding this recall call 949-453-4000, ext 1423.
Quantity in commerce
62,063 units total (14,420 units USA)
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to: ARGENTINA AUSTRALIA BANGLADESH BELGIUM CANADA CHILE CHINA FRANCE GERMANY GREAT BRITAIN GREECE HUNGARY INDIA INDONESIA IRAQ ISRAEL ITALY JAPAN LIBYA MALAYSIA NEPAL NETHERLANDS NEW ZEALAND PERU POLAND PORTUGAL QATAR ROMANIA RUSSIA RWANDA SAUDI ARABIA SOUTH AFRICA SPAIN SRI LANKA SWITZERLAND TAIWAN THAILAND TRINIDAD & TOBAGO TURKEY UNITED ARAB EMIRATES
Product codes and lots
Nasal CPAP Prongs (10 pack), Lot Numbers 13060603XX through 14032503XX; 13082003XX through 14032503XX; 13091903XX through 14032503XX; 13082603XX through 14032503XX; 13091903XX through 14032503XX; 14022603XX through 14032503XX; 14022603XX through 14032503XX; 14031203XX through 14032503XX; 14031203XX through 14032503XX; 14031203XX through 14032503XX; 14031203XX through 14032503XX. Bubble CPAP Starter Kits, Lot Numbers 131007 through 140325.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1539-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1539-2014Class ITerminatedVoluntary: Firm initiated