Z-2735-2014Class II

Verigene BC-GP Kit (Catalog number 20-005-018); 1) 20 Verigene BC-GP Test Cartridges; are qualitative...

Luminex Corporation / initiated 2013-06-17 / Terminated / root cause: packaging

Nanosphere, Inc. began a recall of Verigene BC-GP Kit (Catalog number 20-005-018); 1) 20 Verigene BC-GP Test Cartridges; are qualitative multiplexed in vitro diagnostic tests for simultaneous... on . FDA classified it as Class II and gives the reason as "There is a specific Extraction Tray lot containing Tips that may slightly increase the occurrence of Process SP Processing Error when used due to the Processor SP experiencing difficulty releasing the Tip to the final...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2735-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Luminex Corporation
Product
Verigene BC-GP Kit (Catalog number 20-005-018); 1) 20 Verigene BC-GP Test Cartridges; are qualitative multiplexed in vitro diagnostic tests for simultaneous detection and identification of selected gram-positive (BC-GP) and gram-negative (BC-GN) bacteria and resistance markers.
Product code
PAM Gram-Positive Bacteria And Their Resistance Markers
Reason for recall
There is a specific Extraction Tray lot containing Tips that may slightly increase the occurrence of Process SP Processing Error when used due to the Processor SP experiencing difficulty releasing the Tip to the final location in the Tip Assembly. The difficult releasing the Tip can result in a Processing Error and occasionally the Tip to be dropped into the Processor SP.
Root cause tag
packaging
FDA root cause
Process change control
Action
Undated recall notification letters were sent to all direct accounts (customers) June 17 - 19, 2013. The letters included instructions for customers to: 1) check their inventory for the recalled products; 2) set any recalled products aside for return; and, 3) contact Nanosphere to obtain replacement tip holders. Customers with questions about this recall can contact Nanosphere via e-mail at [email protected] or by calling 888-837-4436.
Quantity in commerce
1,660 extraction trays
Distribution
Worldwide Distribution:US (Nationwide) including states of: AZ, CA, FL, IL, KY, MD, MN, MT, OH, PA, TN, TX, VA, WA, WI; and countries of: Belgium, China, and Saudi Arabia.
Product codes and lots
Part Number: 20-009-018; Lot Numbers: 051713018B, 052413018C

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2735-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2735-2014Class IITerminatedVoluntary: Firm initiated
Z-2736-2014Class IITerminatedVoluntary: Firm initiated