Z-2028-2014Class II
Sarns Disposable Centrifugal Pump without X-Coating,164275, is a sterile, single use device. For use in...
Terumo Cardiovascular Systems Corporation / initiated 2014-04-14 / Terminated / root cause: design
Terumo Cardiovascular Systems Corporation began a recall of Sarns Disposable Centrifugal Pump without X-Coating,164275, is a sterile, single use device. For use in cardiopulmonary bypass procedures on . FDA classified it as Class II and gives the reason as "During set-up and priming of the bypass circuit, leaks were detected on the centrifugal pumphead". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2028-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Sarns Disposable Centrifugal Pump without X-Coating,164275, is a sterile, single use device. For use in cardiopulmonary bypass procedures.
- Product code
- KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type
- Reason for recall
- During set-up and priming of the bypass circuit, leaks were detected on the centrifugal pumphead.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Urgent Medical Device Recall letters, dated April 11, 2014, were sent to customers. The letters identified the affected product, as well as the reason for the correction, the potential hazard, the correction, the affected population, and instructions for customers. Customers are to review the notice and ensure that all users have received a copy. Additionally, customers are to confirm receipt of the communication by returning the attached Customer Response Form as instructed; and, return all effected centrifugal pumps. Terumo CVS will issue a Returned Goods Authorization upon receipt of the Response Form. Questions should be directed to Terumo CVS Customer Service at 800-521-2818.
- Distribution
- Nationwide Distribution
- Product codes and lots
- Lot No. RA07, Lot No. QP02
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2028-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2029-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2030-2014 | Class II | Terminated | Voluntary: Firm initiated |