Z-2029-2014Class II

Sarns Disposable Centrifugal Pump with X-Coating, 164275X, is a sterile, single use device. For use in...

Terumo Cardiovascular Systems Corporation / initiated 2014-04-14 / Terminated / root cause: design

Terumo Cardiovascular Systems Corporation began a recall of Sarns Disposable Centrifugal Pump with X-Coating, 164275X, is a sterile, single use device. For use in cardiopulmonary bypass procedures on . FDA classified it as Class II and gives the reason as "During set-up and priming of the bypass circuit, leaks were detected on the centrifugal pumphead". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2029-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Sarns Disposable Centrifugal Pump with X-Coating, 164275X, is a sterile, single use device. For use in cardiopulmonary bypass procedures.
Product code
KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type
Reason for recall
During set-up and priming of the bypass circuit, leaks were detected on the centrifugal pumphead.
Root cause tag
design
FDA root cause
Device Design
Action
Urgent Medical Device Recall letters, dated April 11, 2014, were sent to customers. The letters identified the affected product, as well as the reason for the correction, the potential hazard, the correction, the affected population, and instructions for customers. Customers are to review the notice and ensure that all users have received a copy. Additionally, customers are to confirm receipt of the communication by returning the attached Customer Response Form as instructed; and, return all effected centrifugal pumps. Terumo CVS will issue a Returned Goods Authorization upon receipt of the Response Form. Questions should be directed to Terumo CVS Customer Service at 800-521-2818.
Distribution
Nationwide Distribution
Product codes and lots
Lot No. RC07, QM03, QN07, and RA09.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2029-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2028-2014Class IITerminatedVoluntary: Firm initiated
Z-2029-2014Class IITerminatedVoluntary: Firm initiated
Z-2030-2014Class IITerminatedVoluntary: Firm initiated