510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1489-2014Class II

ev3 EverCross 0.035" OTW PTA Dilatation Catheter, 6mm x 120mm x 135cm, REF: AB35W06120135. and ev3 EverCross...

Medtronic, Inc. / initiated 2014-04-09 / Terminated / root cause: labeling and UDI

Ev3, Inc. began a recall of ev3 EverCross 0.035" OTW PTA Dilatation Catheter, 6mm x 120mm x 135cm, REF: AB35W06120135. and ev3 EverCross 0.035" OTW PTA Dilatation Catheter, 6mm x 40mm x... on . FDA classified it as Class II and gives the reason as "Covidien is recalling 2 production lots of EverCross 0.035 PTA Balloon Catheter because of a product labeling error. There is a discrepancy in the PTA Balloon length and the delivery catheter length printed on the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1489-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
ev3 EverCross 0.035" OTW PTA Dilatation Catheter, 6mm x 120mm x 135cm, REF: AB35W06120135. and ev3 EverCross 0.035" OTW PTA Dilatation Catheter, 6mm x 40mm x, 80cm REF: AB35W06040080. Peripheral Dilatation Catheter
Product code
DQY Catheter, Percutaneous
Reason for recall
Covidien is recalling 2 production lots of EverCross 0.035 PTA Balloon Catheter because of a product labeling error. There is a discrepancy in the PTA Balloon length and the delivery catheter length printed on the product labeling.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
Consignees were sent a Covidien "Urgent Product Recall" letter dated April 09, 2014. The letter described the problem and the product involved in the recall. The letter described the required actions which included to immediately quarantine and discontinue use of the affected devices and to return affected product. Requested consignees to complete the Verification Form and return it to their Covidien Sales Representative. For questions they can contact Covidien Customer Service at 1-800-716-6000.
Quantity in commerce
39 units (31 OUS, 8 US)
Distribution
Worldwide Distribution-USA including the states of WV, TX, OK, IN, NY, TX, and NC, and the countries of AUSTRALIA, ISRAEL, TURKEY, UNITED KINGDOM, ITALY, FRANCE, and CHINA.
Product codes and lots
AB35W06120135 Lot: 9855406 Use by: 2017-01-16 AB35W06040080 Lot: 9852591 Use by: 2017-01-16

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1489-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1489-2014Class IITerminatedVoluntary: Firm initiated