510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1486-2014Class II

Merit Laureate 0.035" (0.89mm) Hydrophilic Guide Wire, Angled Tip, Standard Shaft, Catalog No. LWSTDA35260EX...

Merit Medical Systems, Inc. / initiated 2014-04-14 / Terminated / root cause: manufacturing process

Merit Medical Systems, Inc. began a recall of Merit Laureate 0.035" (0.89mm) Hydrophilic Guide Wire, Angled Tip, Standard Shaft, Catalog No. LWSTDA35260EX, 260 cm (102") on . FDA classified it as Class II and gives the reason as "Merit Medical Systems, Inc. is voluntarily conducting a recall due to a discrepancy between the carton and unit labeling for the Merit Laureate Hydrophilic 0.035" 260 cm Guide Wire, catalog number LWSTFS35260EX, lot...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1486-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merit Medical Systems, Inc.
Product
Merit Laureate 0.035" (0.89mm) Hydrophilic Guide Wire, Angled Tip, Standard Shaft, Catalog No. LWSTDA35260EX, 260 cm (102").
Product code
DQX Wire, Guide, Catheter
Reason for recall
Merit Medical Systems, Inc. is voluntarily conducting a recall due to a discrepancy between the carton and unit labeling for the Merit Laureate Hydrophilic 0.035" 260 cm Guide Wire, catalog number LWSTFS35260EX, lot number K556399 (173 units), which has a straight tip and a stiff shaft. Lot K556399 was partially mixed with another lot of Laureate 0.035" 260 cm Guide Wire, catalog number LWSTDA3
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Customers were notified on 4/14/14 via certified letter with instructions to identify product, discontinue use, and return products.
Quantity in commerce
186
Distribution
Worldwide Distribution: USA Nationwide and China. No military/govt/VA consignees.
Product codes and lots
Lot No. K556402, Expires 2016-10

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1486-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1485-2014Class IITerminatedVoluntary: Firm initiated
Z-1486-2014Class IITerminatedVoluntary: Firm initiated