Merit Laureate 0.035" (0.89mm) Hydrophilic Guide Wire, Angled Tip, Standard Shaft, Catalog No. LWSTDA35260EX...
Merit Medical Systems, Inc. began a recall of Merit Laureate 0.035" (0.89mm) Hydrophilic Guide Wire, Angled Tip, Standard Shaft, Catalog No. LWSTDA35260EX, 260 cm (102") on . FDA classified it as Class II and gives the reason as "Merit Medical Systems, Inc. is voluntarily conducting a recall due to a discrepancy between the carton and unit labeling for the Merit Laureate Hydrophilic 0.035" 260 cm Guide Wire, catalog number LWSTFS35260EX, lot...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1486-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merit Medical Systems, Inc.
- Product
- Merit Laureate 0.035" (0.89mm) Hydrophilic Guide Wire, Angled Tip, Standard Shaft, Catalog No. LWSTDA35260EX, 260 cm (102").
- Product code
- DQX Wire, Guide, Catheter
- Reason for recall
- Merit Medical Systems, Inc. is voluntarily conducting a recall due to a discrepancy between the carton and unit labeling for the Merit Laureate Hydrophilic 0.035" 260 cm Guide Wire, catalog number LWSTFS35260EX, lot number K556399 (173 units), which has a straight tip and a stiff shaft. Lot K556399 was partially mixed with another lot of Laureate 0.035" 260 cm Guide Wire, catalog number LWSTDA3
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Customers were notified on 4/14/14 via certified letter with instructions to identify product, discontinue use, and return products.
- Quantity in commerce
- 186
- Distribution
- Worldwide Distribution: USA Nationwide and China. No military/govt/VA consignees.
- Product codes and lots
- Lot No. K556402, Expires 2016-10
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1485-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1486-2014 | Class II | Terminated | Voluntary: Firm initiated |