510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1636-2014Class II

TRIGEN INTERTAN NAIL: (1) REF 71675201, 10 MM X 18 CM, 125 degree, (2) REF 71675202, 11.5 MM X 18 CM, 125...

Smith & Nephew, Inc. / initiated 2014-04-04 / Terminated / root cause: manufacturing process

Smith & Nephew Inc began a recall of TRIGEN INTERTAN NAIL: (1) REF 71675201, 10 MM X 18 CM, 125 degree, (2) REF 71675202, 11.5 MM X 18 CM, 125 degree, (3) REF 71675203, 13 MM X 18 CM, 125 degree... on . FDA classified it as Class II and gives the reason as "A manufacturing error resulted in the set screw being inserted too deep in the affected nails". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1636-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
TRIGEN INTERTAN NAIL: (1) REF 71675201, 10 MM X 18 CM, 125 degree, (2) REF 71675202, 11.5 MM X 18 CM, 125 degree, (3) REF 71675203, 13 MM X 18 CM, 125 degree, (4) REF 71675204, 10 MM X 20 CM, 125 degree, (5) REF 71675205, 11.5 MM X 20 CM, 125 degree, (6) REF 71675207, 10 MM X 18 CM, 130 degree, (7) REF 71675208, 11.5 MM X 18 CM, 130 degree, (8) REF 71675211, 11.5 MM X 20 CM, 130 degree, Smith & Nephew, Inc. Orthopedic fixation device.
Product code
JDS Nail, Fixation, Bone
Reason for recall
A manufacturing error resulted in the set screw being inserted too deep in the affected nails.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm initiated the recall via an "Urgent - Product Recall 1st Notification - Urgent" letter and e-mail on 04/04/2014, identifying the product, explaining the problem, and the potential risk associated with using the product. Customers are to inspect their inventory, and locate and quarantine any affected product. Additionally, the Inventory Return Certification form should be completed and returned along with any affected product identified, once a Return Authorization number has been obtained from Smith & Nephew's Safety Affairs Department.
Quantity in commerce
1,253 units total
Distribution
Worldwide Distribution -- USA, France, Germany, Sweden, Switzerland, UK, Canada, Australia, Japan, China, India, Dubai, South Africa, South Korea, Mexico, and Venezuela.
Product codes and lots
TRIGEN INTERTAN NAIL:  (1) REF 71675201, Lot Numbers: 13LT36305, 13LT36306, 13LT36547, 13LT36548, 13LT37009, 13LT37010, 13MT37347, 13MT37665, 13MT37667, 13MT37668, 14AT38179, 14AT38180, 14AT38181, 14T38182, 14AT38183, 14AT38189, 14AT38190   (2) REF 71675202, Lot Numbers: 13LT36308, 13LT36549, 13LT37011, 13MT37348, 14AT38750  (3) REF 71675203, Lot Numbers: 13KT36106, 13LT36596, 14AT38494  (4) REF 71675204, Lot Numbers: 13LT36597, 13LT37012, 14AT38495, 14AT38660  (5) REF 71675205, Lot Number 13LT36598  (6) REF 71675207, Lot Numbers: 13LT363089, 13LT36310, 13LT36599, 13LT37058, 13LT37059, 13LT37063, 13MT37264, 13MT37265, 13MT37266, 13MT37329, 13MT37349, 13MT37350, 13MT37669, 13MT37671, 13MT37672  (7) REF 71675208, Lot Numbers: 13LT36311, 13LT36312, 13LT36600, 13LT37060, 13LT37061, 13MT37351, 13MT37666, 13MT37963  (8) REF 71675211, Lot Number 13LT36313

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1636-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1636-2014Class IITerminatedVoluntary: Firm initiated
Z-1637-2014Class IITerminatedVoluntary: Firm initiated