Z-1544-2014Class II

12-Instrument Sterilization Tray

Stryker Corp. / initiated 2014-04-16 / Terminated / root cause: sterility

Stryker Endoscopy began a recall of 12-Instrument Sterilization Tray; Product Usage: The 12-Instrument Tray is used to hold and protect surgical instruments during the sterilization process... on . FDA classified it as Class II and gives the reason as "Gravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-1544-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
12-Instrument Sterilization Tray; Product Usage: The 12-Instrument Tray is used to hold and protect surgical instruments during the sterilization process. Specifically, this tray is used for the Conquest Manual Instrument line. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held. The tray features silicon device holders that secure devices during the sterilization process.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Gravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing.
Root cause tag
sterility
FDA root cause
Reprocessing Controls
Action
Stryker sent an Urgent Device Corrections letters dated April 18, 2014 to all US customers via certified mail. Stryker Sales representative and Stryker international distributors were notified with the letter and response card by e-mail. International sites are responsible for notifying customers. The letter identified the affected product, problem, potential hazards and actions to be taken. Customers were instructed to complete and sign the enclosed Acknowledgement of Receipt form and scan and e-mail to [email protected]. For questions call 1-800-624-4422 option 3 or via email at [email protected].
Quantity in commerce
2591 trays
Distribution
Worldwide distribution
Product codes and lots
Model number: 0242000012; Lot/serial numbers: 021464-2, 39824-1-1, 4014, 643893-1-4, 27419, 42901-1-1, 35728-2-1,  48639-1-8, 021464-3, 40288-1-1, 43020, 43893-1-6, 27701, 43104-1-1,  36175-1-1, 48956-1-1, 23027, 40629-3-,1, 43622, 43893-1-7, 29129,  43270-1-1, 36421-1-1, 48956-1-1A, 23398, 40721-1-1, 50476,  43893-1-8, 29670, 43523-1-1, 36424-1-1, 51304-1-1, 23542,  41231-4-1, 54990, 43893-1-9, 30250, 43523-1-1A, 37251-1-1,  51304-1-1A, 23846, 41478-1-1, 7J1, 48639-1-2, 30919, 43674-1-1,  37392-1-1, 52123-1-1, 24073, 41745-1-1, 123, 48639-1-3, 3218., 43893-1-1, 37794-1-1, 52956-1-1, 24511, 42086-1-1, 28646-1-1,  48639-1-4, 33226, 43893-1-10, 38867-1-1, 54346-1-1, 26445,  42306-1-1, 29005-1-1, 48639-1-5, 33961, 43893-1-1, 1SS15327, 26843,  42526-1-1, 30384-1-1, 48639-1-7, 35933, 43893-1-12, 37991, 36610,  43893-1-13, 39219, 43893-1-2, 43893-1-14.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1544-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1544-2014Class IITerminatedVoluntary: Firm initiated