12-Instrument Sterilization Tray
Stryker Endoscopy began a recall of 12-Instrument Sterilization Tray; Product Usage: The 12-Instrument Tray is used to hold and protect surgical instruments during the sterilization process... on . FDA classified it as Class II and gives the reason as "Gravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-1544-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- 12-Instrument Sterilization Tray; Product Usage: The 12-Instrument Tray is used to hold and protect surgical instruments during the sterilization process. Specifically, this tray is used for the Conquest Manual Instrument line. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held. The tray features silicon device holders that secure devices during the sterilization process.
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Gravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing.
- Root cause tag
- sterility
- FDA root cause
- Reprocessing Controls
- Action
- Stryker sent an Urgent Device Corrections letters dated April 18, 2014 to all US customers via certified mail. Stryker Sales representative and Stryker international distributors were notified with the letter and response card by e-mail. International sites are responsible for notifying customers. The letter identified the affected product, problem, potential hazards and actions to be taken. Customers were instructed to complete and sign the enclosed Acknowledgement of Receipt form and scan and e-mail to [email protected]. For questions call 1-800-624-4422 option 3 or via email at [email protected].
- Quantity in commerce
- 2591 trays
- Distribution
- Worldwide distribution
- Product codes and lots
- Model number: 0242000012; Lot/serial numbers: 021464-2, 39824-1-1, 4014, 643893-1-4, 27419, 42901-1-1, 35728-2-1, 48639-1-8, 021464-3, 40288-1-1, 43020, 43893-1-6, 27701, 43104-1-1, 36175-1-1, 48956-1-1, 23027, 40629-3-,1, 43622, 43893-1-7, 29129, 43270-1-1, 36421-1-1, 48956-1-1A, 23398, 40721-1-1, 50476, 43893-1-8, 29670, 43523-1-1, 36424-1-1, 51304-1-1, 23542, 41231-4-1, 54990, 43893-1-9, 30250, 43523-1-1A, 37251-1-1, 51304-1-1A, 23846, 41478-1-1, 7J1, 48639-1-2, 30919, 43674-1-1, 37392-1-1, 52123-1-1, 24073, 41745-1-1, 123, 48639-1-3, 3218., 43893-1-1, 37794-1-1, 52956-1-1, 24511, 42086-1-1, 28646-1-1, 48639-1-4, 33226, 43893-1-10, 38867-1-1, 54346-1-1, 26445, 42306-1-1, 29005-1-1, 48639-1-5, 33961, 43893-1-1, 1SS15327, 26843, 42526-1-1, 30384-1-1, 48639-1-7, 35933, 43893-1-12, 37991, 36610, 43893-1-13, 39219, 43893-1-2, 43893-1-14.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1544-2014 | Class II | Terminated | Voluntary: Firm initiated |