Z-1698-2014Class II

Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (NCL-L-GSTpi-438), a liquid tissue...

Leica Microsystems, Inc. / initiated 2014-02-26 / Terminated / root cause: labeling and UDI

Leica Microsystems, Inc. began a recall of Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (NCL-L-GSTpi-438), a liquid tissue culture supernatant. For in vitro diagnostic use on . FDA classified it as Class II and gives the reason as "Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (Product Code: NCL-L-GSTpi-438) does not function as intended up to the expiry date on the product labeling. This was detected through an ongoing...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1698-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Leica Microsystems, Inc.
Product
Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (NCL-L-GSTpi-438), a liquid tissue culture supernatant. For in vitro diagnostic use.
Product code
NJT Immunohistochemistry Reagents And Kits
Reason for recall
Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (Product Code: NCL-L-GSTpi-438) does not function as intended up to the expiry date on the product labeling. This was detected through an ongoing stability program. There is a linkage between the age of the product and staining intensity.
Root cause tag
labeling and UDI
FDA root cause
Incorrect or no expiration date
Action
Leica Biosystems sent a Medical Device Recall Notification letter dated February 26, 2014 to all direct accounts. The letter advised that clinicians may want to review results obtained from the affected products. The letter also included instructions for customers to: 1) appropriately destroy any unused or partially used NCL-L-GSTpi-438 (lots 6018777, 6023260,6026297, 6026980) reagent or confirm that they are not in stock; 2) complete and return the attached Medical Device Recall Notification Acknowledgement Form to 847-236-3747; and, 3) contact the local Leica representative immediately if replacement is required. Customers with questions or concerns about this recall can contact Leica via e-mail at [email protected]. For questions regarding this recall call 847-405-5413.
Quantity in commerce
33 units
Distribution
Nationwide Distribution including AZ, CA, FL, and NY.
Product codes and lots
Product Code: NCL-L-GSTpi-438; Lot Numbers and Expiration Dates: 6018777, Expires 2014-05; 6023260, Expires 2014-12; 6026297, Expires 2015-05; 6026980, Expires 2015-06

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1698-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1698-2014Class IITerminatedVoluntary: Firm initiated