Panda Freestanding Infant Warmers, integrated with Nellcor SpO2. Controlled infrared heat to neonates who are...
GE Healthcare began a recall of Panda Freestanding Infant Warmers, integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own... on . FDA classified it as Class II and gives the reason as "Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1549-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Panda Freestanding Infant Warmers, integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monitoring feature used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by a SpO2 sensor).
- Product code
- FMT Warmer, Infant Radiant
- Reason for recall
- Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- GE Healthcare notified Biomedical/Clinical Engineering managers, Labor & Delivery/ NICU Nurse managers, Respiratory Therapy Managers and Directors of Risk Management by letter 04/02/2014. End users were advised that they may continue to use the system provided they follow the GE recommended actions: 1) If in use with a patient, turn off the Nellcor SpO2 function in the control panel and switch to an approved alternate form of SpO2 measurement. 2) Prior to using the affected GE warmer with the Nellcor SpO2 function, or resuming the use of this function with any patient, contact your hospital Bio-Med or qualified technician to check the Nellcor SpO2 alarm functionality using either an adult Nellcor SpO2 sensor or a Nellcor SpO2 simulator following the applicable instructions in the Appendix 3) If the test activates a Low Pulse Rate alarm, the user-adjusted Nellcor SpO2 alarm limits are functioning properly and the warmer can be placed back in clinical service. End users were also advised that alarm functionality must be re-checked following every service event requiring access to the internal electronics and during annual preventative maintenance, until revised software is installed. 4) If the test does NOT activate a Low Pulse Rate alarm, remove the warmer from clinical service and contact your local GE Healthcare Service Representative.
- Quantity in commerce
- 21 units
- Distribution
- Worldwide Distribution.
- Product codes and lots
- HDJS51142 HDJS52094 HDJS52095 HDJS52153 HDJS52154 HDJS52155 HDJS52156 HDJS52223 HDJS52224 HDJS52517 HDJS52518 HDJS52519 HDJS52520 HDJS52521 HDJS52522 HDJS52974 HDJS52975 HDJS52976 HDJS52977 HDJS52978 HDJS52979
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1547-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1548-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1549-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1550-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1551-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.