Arrow Percutaneous Sheath Introducer Kits The Percutaneous sheath introducer products permit venous or...
Arrow International Inc began a recall of Arrow Percutaneous Sheath Introducer Kits The Percutaneous sheath introducer products permit venous or arterial access and device introduction to the central... on . FDA classified it as Class II and gives the reason as "Arrow is recalling the Arrow Percutaneous Sheath Introducer Kits due to dilator tip damage noted during internal inspection. Dilator tip damage has the potential to result in vessel damage". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1713-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arrow International LLC (subsidiary of Teleflex Incorporated)
- Product
- Arrow Percutaneous Sheath Introducer Kits The Percutaneous sheath introducer products permit venous or arterial access and device introduction to the central circulation.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Arrow is recalling the Arrow Percutaneous Sheath Introducer Kits due to dilator tip damage noted during internal inspection. Dilator tip damage has the potential to result in vessel damage.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Teleflex/Arrow International sent an Urgent Medical Device Recall Notification letter dated April 17, 2014, to distributors and end users. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue use and quarantine any affected product and to complete the enclosed Recall Acknowledgement Form and fax it to 1-800-343-2935 to return product. Customers would be contacted by a customer service representative for a Return Goods Authorization (RGA) Numbers and would be provided instructions for the return of the product. Customers should complete the form whether they have affected product or not. Customers with questions were instructed to call 1-800-343-2935. For questions regarding this recall call 610-378-0131.
- Quantity in commerce
- 15424
- Distribution
- Worldwide Distribution - USA (nationwide) and Canada.
- Product codes and lots
- Product codes and lot numbers: AK-09903-A 23F13G0162 AK-09903-S RF3064511 CDC-09903-1A 23F13G0642 AK-09903-J 23F13G0400 ASK-09903-CMC RF3064673 NL-09903-S 23F13G0630 AK-09903-LFSP RF3065449 ASK-09903-CMC 23F13G0343 NR-09903-S 23F13G0180 AK-09903-S 23F13G0204 ASK-09903-NKC 23F13G0336 SI-09903-E RF3064710 AK-09903-S RF3064956 ASK-09903-TJ 23F13G0567 SI-09903-E RF3064915 AK-09903-S 23F13H0195 CDC-09903-1A RF3065316
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1713-2014 | Class II | Terminated | Voluntary: Firm initiated |