Z-0179-2015Class II

Thyroglobulin (1D4) is a mouse anti-human monoclonal antibody produced as a tissue culture supernatant, and...

Leica Microsystems, Inc. / initiated 2014-04-18 / Terminated / root cause: labeling and UDI

Leica Microsystems, Inc. began a recall of Thyroglobulin (1D4) is a mouse anti-human monoclonal antibody produced as a tissue culture supernatant, and supplied in Tris buffered saline with carrier... on . FDA classified it as Class II and gives the reason as "Bond Ready to Use Primary Antibody Thyroglobulin does not function as intended up to the expiry date on the product labelling. This was detected through an on-going stability program. There is a link between the age of...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0179-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Leica Microsystems, Inc.
Product
Thyroglobulin (1D4) is a mouse anti-human monoclonal antibody produced as a tissue culture supernatant, and supplied in Tris buffered saline with carrier protein, containing 0.35% ProClin 950 as a preservative. Total volume = 7 mL. This reagent is for in vitro diagnostic use. Thyroglobulin (1D4) monoclonal antibody is intended to be used for the qualitative identification by light microscopy of thyroglobulin in formalin-fixed, paraffin-embedded tissue by immunohistochemical staining using an automated Bond system. The clinical interpretation of any staining or its absence should be complemented by morphological studies and proper controls and should be evaluated within the context of the patient's clinical history and other diagnostic tests by a qualified pathologist.
Product code
NJT Immunohistochemistry Reagents And Kits
Reason for recall
Bond Ready to Use Primary Antibody Thyroglobulin does not function as intended up to the expiry date on the product labelling. This was detected through an on-going stability program. There is a link between the age of the product and staining intensity.
Root cause tag
labeling and UDI
FDA root cause
Unknown/Undetermined by firm
Action
Leica Biosystems sent a Medical Device Recall Notification letter dated April 18, 2014, to all affected customers. The letter included instructions for customers to: 1) assess any results previously obtained by using the recalled products; 2) identify and destroy any unused or partly used recalled products or confirm that the recalled products are no longer in stock; and, 3) complete and return the attached Medical Device Recall Notification Acknowledgement Form. Customers with questions were instructed to contact Leica Microsystems by e-mail at [email protected]. For questions regarding this recall call 847-405-5413.
Quantity in commerce
180 units
Distribution
Worldwide Distribution - USA including AR, AZ, CA, CO, CT, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, MT, MY, NC, NH, NJ, NY, OH, ON, OR, PA, TN, TX, UT, VA, WA, WI and the District of Columbia, Internationally to Belgium, Bulgaria, France, Italy, Pakistan, Philippines, Russia, Saudi Arabia, Spain, Turkey, and the United Kingdom.
Product codes and lots
Product Code: PA0025; Lot Numbers and Expiration Dates: 24963, Expires 17 DEC 2014; 24228, Expires 15 OCT 2014; 23882, Expires 10 SEP 2014; 23366, Expires 23 JUL 2014

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0179-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0179-2015Class IITerminatedVoluntary: Firm initiated