Z-1514-2014Class II

HexaPOD evo RT Couchtop. To support and aid in positioning a patient during radiation therapy

Elekta Limited / initiated 2014-04-17 / Terminated / root cause: design

Elekta, Inc. began a recall of HexaPOD evo RT Couchtop. To support and aid in positioning a patient during radiation therapy on . FDA classified it as Class II and gives the reason as "If screws are not properly tightened with a torque wrench they may become loose over time". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1514-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
HexaPOD evo RT Couchtop. To support and aid in positioning a patient during radiation therapy.
Product code
JAI Couch, Radiation Therapy, Powered
Reason for recall
If screws are not properly tightened with a torque wrench they may become loose over time.
Root cause tag
design
FDA root cause
Component design/selection
Action
Corrective Action #1 - Important Field Safety Notice 618 01 302 002 was issued April 2014. The Notice informs end users about the problem and the potential clinical impact. It also give a recommendation how the risk can be mitigated - inspect the screws and if loose contact Elekta Service to tighten the screws with 20 Nm. The notice also advises the customer that a modification kit will be released that introduces an improved screw solution. The solution will be implemented on all affected products when available. Corrective Action #2 Permanent Solution - Replace Current Screws - An improved screw has been designed and will be available for Elekta Field Service organization to replace on all affected products at the end of April. The new design solution provides a very long used thread depth, exceeding the standard recommended thread-depths by far. This design will prevent the screws from loosening.
Quantity in commerce
266
Distribution
Worldwide Distribution-USA including DC and the states of AZ, CA, CT, FL, GA, IL, HI, KY, MD, MA, MI, MS, MO, NJ, NY, NH, NC, ND, OH, OK, OR, PA, SC, TX, VA, WA, WI, and WY, and the countries of Algeria, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kazakhstan, Mexico, Netherlands, New Zealand, Peru, Poland, Portugal, Russia, Saudi Arabia, South Korea, Span, Sweden, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.
Product codes and lots
Serial Numbers 42188, 42246, 42106, 42153, 42107, 42068, 42012, 42013, 42062, 42009, 42224, 42126, 3005, 42222, 3019, 3000, 42058, 42059, 42057, 42187, 42129, 42159, 3022, 3017, 42189, 42213, 42202, 42212, 42004, 4220, 42113, 42178, 42216, 42207, 2982, 2986, 2991, 42102, 42000, 42032, 42145, 42174, 42008, 42048, 42064, 42070, 42074, 42084, 42115, 42134, 42118, 42152, 42137, 42146, 3008, 2987, 3010, 3018, 42069, 42036, 3029, 42162, 42139, 42011, 42007, 42173, 42109, 42193, 3006, 42053, 42015, 42105, 42124, 42085, 42065, 42094, 42198, 2998, 3003, 3033, 3024, 2980, 42121, 2999, 42027, 42061, 42075, 42079, 42214, 3020, 42192, 42247, 42202, 42155, 42025, 42021, 42127, 42151, 42239, 42033, 42035, 42089, 42204, 42200, 2995, 3030, 42266, 42197, 42077, 42001, 42037, 42147, 42186, 42016, 42028, 42029, 42050, 42042, 42054, 42047, 42051, 42055, 42049, 42262, 42080, 42076, 42063, 42103, 42087, 42117, 42083, 42166, 42101, 42227, 42150, 42156, 42163, 42179, 42180, 42140, 42165, 42157,42234, 42169, 42182, 42183, 42236, 42195, 42223, 42256, 42276, 42260, 42231, 42250, 42243, 42240, 42272, 42235, 42249, 42254, 42245, 42265, 42268, 42267, 42253, 42264, 2993, 2997, 3004, 3031, 3002, 42226, 42258, 42114, 42081, 42090, 42093, 42133, 42181, 42039, 42116, 3016, 29677, 42010, 42073, 42176, 42164, 42031, 3028, 42041, 42038, 42022, 42034, 42088, 42104, 2976, 2979, 2988, 3011, 3015, 3021, 3027, 42238, 42018, 42135, 42252, 42185, 42273, 42060, 42024, 42006, 42082, 42194, 42263, 42232, 42199, 42170, 42161, 42078, 42158, 2985, 2994, 42120, 3032, 42056, 2969, 3034, 2975, 2972, 2984, 42191, 42225, 42228, 42023, 42143, 42122, 42208, 42230, 42211, 42026, 42206, 42066, 42052, 42014, 42218, 42030, 42040, 42244, 42043, 42044, 42045, 42046, 42123, 42255, 42141, 42125, 42072, 42112, 42210, 42017, 420196, 42020, 42067, 42071, 42097, 42100, 42131, 42144, 42136, 42148, 42130, 42111, 42067, 42171, 42203, 4210, 42205, 42096, 42154, 42241, 42229, 2981, 2989, 2971, 2990, 3007, 3013, 2992, 2973, 2978, 30098, 2974, 3014, 3012, 3025, 2996, 3026, 42274

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1514-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1514-2014Class IITerminatedVoluntary: Firm initiated