Z-1514-2014Class II
HexaPOD evo RT Couchtop. To support and aid in positioning a patient during radiation therapy
Elekta Limited / initiated 2014-04-17 / Terminated / root cause: design
Elekta, Inc. began a recall of HexaPOD evo RT Couchtop. To support and aid in positioning a patient during radiation therapy on . FDA classified it as Class II and gives the reason as "If screws are not properly tightened with a torque wrench they may become loose over time". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1514-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- HexaPOD evo RT Couchtop. To support and aid in positioning a patient during radiation therapy.
- Product code
- JAI Couch, Radiation Therapy, Powered
- Reason for recall
- If screws are not properly tightened with a torque wrench they may become loose over time.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Corrective Action #1 - Important Field Safety Notice 618 01 302 002 was issued April 2014. The Notice informs end users about the problem and the potential clinical impact. It also give a recommendation how the risk can be mitigated - inspect the screws and if loose contact Elekta Service to tighten the screws with 20 Nm. The notice also advises the customer that a modification kit will be released that introduces an improved screw solution. The solution will be implemented on all affected products when available. Corrective Action #2 Permanent Solution - Replace Current Screws - An improved screw has been designed and will be available for Elekta Field Service organization to replace on all affected products at the end of April. The new design solution provides a very long used thread depth, exceeding the standard recommended thread-depths by far. This design will prevent the screws from loosening.
- Quantity in commerce
- 266
- Distribution
- Worldwide Distribution-USA including DC and the states of AZ, CA, CT, FL, GA, IL, HI, KY, MD, MA, MI, MS, MO, NJ, NY, NH, NC, ND, OH, OK, OR, PA, SC, TX, VA, WA, WI, and WY, and the countries of Algeria, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kazakhstan, Mexico, Netherlands, New Zealand, Peru, Poland, Portugal, Russia, Saudi Arabia, South Korea, Span, Sweden, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.
- Product codes and lots
- Serial Numbers 42188, 42246, 42106, 42153, 42107, 42068, 42012, 42013, 42062, 42009, 42224, 42126, 3005, 42222, 3019, 3000, 42058, 42059, 42057, 42187, 42129, 42159, 3022, 3017, 42189, 42213, 42202, 42212, 42004, 4220, 42113, 42178, 42216, 42207, 2982, 2986, 2991, 42102, 42000, 42032, 42145, 42174, 42008, 42048, 42064, 42070, 42074, 42084, 42115, 42134, 42118, 42152, 42137, 42146, 3008, 2987, 3010, 3018, 42069, 42036, 3029, 42162, 42139, 42011, 42007, 42173, 42109, 42193, 3006, 42053, 42015, 42105, 42124, 42085, 42065, 42094, 42198, 2998, 3003, 3033, 3024, 2980, 42121, 2999, 42027, 42061, 42075, 42079, 42214, 3020, 42192, 42247, 42202, 42155, 42025, 42021, 42127, 42151, 42239, 42033, 42035, 42089, 42204, 42200, 2995, 3030, 42266, 42197, 42077, 42001, 42037, 42147, 42186, 42016, 42028, 42029, 42050, 42042, 42054, 42047, 42051, 42055, 42049, 42262, 42080, 42076, 42063, 42103, 42087, 42117, 42083, 42166, 42101, 42227, 42150, 42156, 42163, 42179, 42180, 42140, 42165, 42157,42234, 42169, 42182, 42183, 42236, 42195, 42223, 42256, 42276, 42260, 42231, 42250, 42243, 42240, 42272, 42235, 42249, 42254, 42245, 42265, 42268, 42267, 42253, 42264, 2993, 2997, 3004, 3031, 3002, 42226, 42258, 42114, 42081, 42090, 42093, 42133, 42181, 42039, 42116, 3016, 29677, 42010, 42073, 42176, 42164, 42031, 3028, 42041, 42038, 42022, 42034, 42088, 42104, 2976, 2979, 2988, 3011, 3015, 3021, 3027, 42238, 42018, 42135, 42252, 42185, 42273, 42060, 42024, 42006, 42082, 42194, 42263, 42232, 42199, 42170, 42161, 42078, 42158, 2985, 2994, 42120, 3032, 42056, 2969, 3034, 2975, 2972, 2984, 42191, 42225, 42228, 42023, 42143, 42122, 42208, 42230, 42211, 42026, 42206, 42066, 42052, 42014, 42218, 42030, 42040, 42244, 42043, 42044, 42045, 42046, 42123, 42255, 42141, 42125, 42072, 42112, 42210, 42017, 420196, 42020, 42067, 42071, 42097, 42100, 42131, 42144, 42136, 42148, 42130, 42111, 42067, 42171, 42203, 4210, 42205, 42096, 42154, 42241, 42229, 2981, 2989, 2971, 2990, 3007, 3013, 2992, 2973, 2978, 30098, 2974, 3014, 3012, 3025, 2996, 3026, 42274
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1514-2014 | Class II | Terminated | Voluntary: Firm initiated |