Liquid Mouse Monoclonal Antibody CD15 (NCL-L-CD15), a liquid tissue culture supernatant containing 15 mM...
Leica Microsystems, Inc. began a recall of Liquid Mouse Monoclonal Antibody CD15 (NCL-L-CD15), a liquid tissue culture supernatant containing 15 mM sodium azide as a preservative. For in vitro... on . FDA classified it as Class II and gives the reason as "Product does not function as intended up to the expiry date on the product labeling. There is a link between the age of the product and staining intensity". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1710-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Leica Microsystems, Inc.
- Product
- Liquid Mouse Monoclonal Antibody CD15 (NCL-L-CD15), a liquid tissue culture supernatant containing 15 mM sodium azide as a preservative. For in vitro diagnostic use.
- Product code
- NJT Immunohistochemistry Reagents And Kits
- Reason for recall
- Product does not function as intended up to the expiry date on the product labeling. There is a link between the age of the product and staining intensity.
- Root cause tag
- labeling and UDI
- FDA root cause
- Incorrect or no expiration date
- Action
- Leica sent an Urgent Field Safety Correction Notice letter dated January 10, 2014 was sent ot all affected customers. The notice included instructions for recipients to: 1) destroy any unused or partially used NCL-L-CD15 (lots 6015073 and 6017936) reagent or confirm that the products are no longer in stock; 2) indicate on the attached Field Safety Correction Notice Acknowledgement Fom which action was taken; and, 3) contact the local Leica represetnative immediately if a replacement is required. Customers with any concerns about this notice can contact Leica via e-mail at [email protected].
- Quantity in commerce
- 145 units
- Distribution
- Worldwide Distribution - US Nationwide in the states of CA, CT FL, IL, MO, MS, NC, NY, PA, and the countries of Argentina, Australia, Brazil, Colombia, Italy, Malaysia, New Zealand, Nigeria, Poland, Romania, Russia, Saudi Arabia, Serbia, South Africa, Sweden, Taiwan, Tunisia, Turkey, and United Kingdom
- Product codes and lots
- Product Code: NCL-L-CD15; Lot Numbers and Expiration Dates: 6017936, Expires 2014-08; 6015073, Expires 2014-03