510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1710-2014Class II

Liquid Mouse Monoclonal Antibody CD15 (NCL-L-CD15), a liquid tissue culture supernatant containing 15 mM...

Leica Microsystems, Inc. / initiated 2014-01-10 / Terminated / root cause: labeling and UDI

Leica Microsystems, Inc. began a recall of Liquid Mouse Monoclonal Antibody CD15 (NCL-L-CD15), a liquid tissue culture supernatant containing 15 mM sodium azide as a preservative. For in vitro... on . FDA classified it as Class II and gives the reason as "Product does not function as intended up to the expiry date on the product labeling. There is a link between the age of the product and staining intensity". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1710-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Leica Microsystems, Inc.
Product
Liquid Mouse Monoclonal Antibody CD15 (NCL-L-CD15), a liquid tissue culture supernatant containing 15 mM sodium azide as a preservative. For in vitro diagnostic use.
Product code
NJT Immunohistochemistry Reagents And Kits
Reason for recall
Product does not function as intended up to the expiry date on the product labeling. There is a link between the age of the product and staining intensity.
Root cause tag
labeling and UDI
FDA root cause
Incorrect or no expiration date
Action
Leica sent an Urgent Field Safety Correction Notice letter dated January 10, 2014 was sent ot all affected customers. The notice included instructions for recipients to: 1) destroy any unused or partially used NCL-L-CD15 (lots 6015073 and 6017936) reagent or confirm that the products are no longer in stock; 2) indicate on the attached Field Safety Correction Notice Acknowledgement Fom which action was taken; and, 3) contact the local Leica represetnative immediately if a replacement is required. Customers with any concerns about this notice can contact Leica via e-mail at [email protected].
Quantity in commerce
145 units
Distribution
Worldwide Distribution - US Nationwide in the states of CA, CT FL, IL, MO, MS, NC, NY, PA, and the countries of Argentina, Australia, Brazil, Colombia, Italy, Malaysia, New Zealand, Nigeria, Poland, Romania, Russia, Saudi Arabia, Serbia, South Africa, Sweden, Taiwan, Tunisia, Turkey, and United Kingdom
Product codes and lots
Product Code: NCL-L-CD15; Lot Numbers and Expiration Dates: 6017936, Expires 2014-08; 6015073, Expires 2014-03

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1710-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1710-2014Class IITerminatedVoluntary: Firm initiated