Recalled product is a component of the Newport Medical Instruments e360 Ventilator: single board computer...
Nellcor Puritan Bennett Inc. (dba Covidien LP) began a recall of Recalled product is a component of the Newport Medical Instruments e360 Ventilator: single board computer, part number SBC2105A. Intended to provide invasive... on . FDA classified it as Class II and gives the reason as "Reports of the audible alarm not sounding when the ventilator is in an alarm state. The ventilator, as well as the visual and remote alarms, continue to operate normally". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1577-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Recalled product is a component of the Newport Medical Instruments e360 Ventilator: single board computer, part number SBC2105A. Intended to provide invasive or noninvasive ventilator support and monitoring for infant, pediatric, and adult patients with respiratory failure or respiratory insufficiency.
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Reports of the audible alarm not sounding when the ventilator is in an alarm state. The ventilator, as well as the visual and remote alarms, continue to operate normally.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Newport Medical Instruments, Inc. (Newport), a wholly-owned subsidiary of Covidien is conducting a voluntary field corrective action via letters to be sent during the week of April 28, 2014. The letter titled "URGENT MEDICAL DEVICE FIELD SAFETY CORRECTIVE ACTION" informed customers of the recall by providing: reason for recall, product description, information on affected serial numbers regardless of software versions, issue identified, recommended actions to help mitigate risk to the patient, actions to eliminate the occurrence of this issue, important safety reminders, and contact information. Customers are to contact Technical Support Department to receive a software update kit, and then follow the instructions provided once the kit has been received. The attached verification and completion form should be completed and faxed.
- Quantity in commerce
- 5,765
- Distribution
- Worldwide Distribution -- USA, including the states of Arizona, California, Colorado, Connecticut, Florida, Georgia, Kentucky, Massachusetts, Maryland, Missouri, Nevada, Pennsylvania, Rhode Island, Texas, Virginia, and Washington; and, the countries of Algeria, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Chile, China, Colombia, Dominican Republic, Ecuador, Egypt, England, France, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Ireland, Israel, Japan, Kazakhstan, Kenya, Lebanon, Liechtenstein, Lithuania, Mexico, Morocco, Namibia, Nepal, Netherlands, Nicaragua, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Republic of Armenia, Republic of KOS, Russia, Saudi Arabia, Singapore, Slovenia, S. Africa, Sri Lanka, Taiwan, Thailand, Tunisia, Turkey, UAE, Uruguay, Vietnam, and Yemen.
- Product codes and lots
- Part Number: SBC2105A. Model Number(s): e360 Ventilator. Serial numbers N11360719000 through N14360426029.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1577-2014 | Class II | Terminated | Voluntary: Firm initiated |