510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1615-2014Class II

ExploR 7x26mm Modular Radial Stem, Catalog number 11-210062. Elbow prosthesis orthopedic implant

Zimmer GmbH / initiated 2014-04-22 / Terminated / root cause: manufacturing process

Biomet, Inc. began a recall of ExploR 7x26mm Modular Radial Stem, Catalog number 11-210062. Elbow prosthesis orthopedic implant on . FDA classified it as Class II and gives the reason as "This lot of the ExploR Modular Radial Stem may be missing the wedge ramp threads. If the wedge ramp threads are missing from the stem, the likelihood of the set screw loosening from the radial stem intraoperatively is...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1615-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
ExploR 7x26mm Modular Radial Stem, Catalog number 11-210062. Elbow prosthesis orthopedic implant.
Product code
KWI Prosthesis, Elbow, Hemi-, Radial, Polymer
Reason for recall
This lot of the ExploR Modular Radial Stem may be missing the wedge ramp threads. If the wedge ramp threads are missing from the stem, the likelihood of the set screw loosening from the radial stem intraoperatively is increased. If the set screw loosens, a revision surgery may be necessary.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Biomet sent an URGENT MEDICAL DEVICE SAFETY ALERT dated April 22, 2014 to affected customers. The letter identified the affected product, problem and actions to be taken. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Quantity in commerce
8 devices
Distribution
US Nationwide Distribution in the states of NY, MD, NC, SC, MI, OR, and NV.
Product codes and lots
Catalog Number: 11-210062 Lot Number Identification: 972230

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1615-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1615-2014Class IITerminatedVoluntary: Firm initiated