Z-1675-2014Class II

Integra Licox Test Kit REF BC10R Rx Only An accessory provided for use with the LCXO2 and AC31 PtO2 Monitors...

Integra LifeSciences Corporation / initiated 2014-04-16 / Terminated / root cause: manufacturing process

Integra LifeSciences Corp. began a recall of Integra Licox Test Kit REF BC10R Rx Only An accessory provided for use with the LCXO2 and AC31 PtO2 Monitors. It provides users with a secondary means to... on . FDA classified it as Class II and gives the reason as "Integra LifeSciences has identified through a small number of complaints that some lot numbers of Licox Test Kits (Catalog Number BC10R) may provide erroneous "calibration test failure" message when they are used with...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1675-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
Integra Licox Test Kit REF BC10R Rx Only An accessory provided for use with the LCXO2 and AC31 PtO2 Monitors. It provides users with a secondary means to verify the monitor's PtO2 state-of-calibration (but does not provide calibration) when needed by providing a specific electrical signal to the monitor.
Product code
GWM Device, Monitoring, Intracranial Pressure
Reason for recall
Integra LifeSciences has identified through a small number of complaints that some lot numbers of Licox Test Kits (Catalog Number BC10R) may provide erroneous "calibration test failure" message when they are used with the Licox PtO2 Monitor (LCSO2, or may provide an erroneous out of range result with the Licox CMP Brain Tissue Oxygen and Temperature Monitor (AC31).
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Integra sent an "Urgent Voluntary Medical Device Recall/Return Response form notifications dated April 16, 2014 to US customers via traceable courrier service and to non-US customers via traceable email or facsimile. The customer notification advises them of the nature of the issue and the potential for adverse patient consequences; advises them to review their inventory and if they have BC10R Test Kit Resistors with an affected lot number to immediately stop using it; requests they identify and report to Integra if they do or do not have any affected BC10R affected lot numbers; and advises them how to return any affected products in their possession. Customers with questions were instructed to call 1-855-532-1723.
Quantity in commerce
49 test probes
Distribution
Worldwide Distribution - USA including Illinois, Florida, Maryland, Massachusetts, North Carolina, Pennsylvania, South Dakota and Internationally to Australia, Austria, Belgium, Chile, Finland, Germany, Great Britain, Greece, Israel, Italy and Russia.
Product codes and lots
Catalog Number BC10R Lot Nos: 051113, 100913, 140513, 161013, 25013

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1675-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1675-2014Class IITerminatedVoluntary: Firm initiated