Z-1719-2014Class II

MACROLYTE Premie Dispersive Electrode, REF 440-2400, CONMED CORPORATION. The device is applied to the patient...

ConMed Corporation / initiated 2014-05-08 / Terminated / root cause: manufacturing process

ConMed Corporation began a recall of MACROLYTE Premie Dispersive Electrode, REF 440-2400, CONMED CORPORATION. The device is applied to the patient during electrosurgical procedures to provide a... on . FDA classified it as Class II and gives the reason as "ConMed Corporation has received complaints regarding sparking, no output and burning at the cord set connection. The complaints have been confirmed by ConMed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1719-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ConMed Corporation
Product
MACROLYTE Premie Dispersive Electrode, REF 440-2400, CONMED CORPORATION. The device is applied to the patient during electrosurgical procedures to provide a path for RF current to leave the patient and return to the generator.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
ConMed Corporation has received complaints regarding sparking, no output and burning at the cord set connection. The complaints have been confirmed by ConMed.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
ConMed sent an Urgent Medical Device Recall letter dated May 8, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Corrective Action: All returned devices will be received into the facility and the lot number and quantity recorded on to an electronic spreadsheet. The returned product will then be stored in a secure quarantined area. At this time, management has not determined as to how the devices will be dispositioned. A protocol(s) will be written to address the destruction of the returned devices. This will be submitted to NYK-DO for approval prior to any action being taken with the returned devices. CAPA 153136 was opened. Fixtures were revised to allow for placement of the foil/gel substrate from the bottom of the foam. This is to maintain consistent dimensions on the product and not allow for float/movement of the substrate from being offset. For further questions please call (315) 624-3024 or fax to (315) 624-3030.
Quantity in commerce
Domestic - 12,600 units, Foreign - 44,280 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Australia, Belgium, Brazil, Canada, Ecuador, El Salvador, Finland, France, Great Britain, Honduras, India, Indonesia, Israel, Italy, Japan, Latvia, Lebanon, Malaysia, Mexico, Panama, Philippines, Portugal, Saudi Arabia, South Africa, Spain, Thailand, Trinidad, United Arab Emirates and Uruguay.
Product codes and lots
Beginning Lot Code 1204275; Ending Lot Code 1401075 (Lot code indicates the date of manufacture = yymmdds, where s is the shift).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1719-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1719-2014Class IITerminatedVoluntary: Firm initiated